Dentsply Sirona
Instructions for CEREC Bluecam and CEREC Omnicam
2 CEREC Omnicam
2.7 High-level disinfection of the mirror sleeve with the HLD set 66 05 120
62 82 839 D3492
D3492.201.02.06.02 10.2018
9
9.
Insert the metal bracket with the metal sleeve.
10. Place the cover on the container and leave the mirror sleeve in the
disinfectant for at least 12 minutes (CIDEX
®
OPA). The mirror sleeve
will not be damaged if it remains in the disinfectant for longer than
the required time, but removing it shortly after the required time is
recommended.