2 CEREC Omnicam
2.5 Pre-cleaning of the mirror sleeve
Dentsply Sirona
Instructions for CEREC Bluecam and CEREC Omnicam
6
2.5
Pre-cleaning of the mirror sleeve
Clean the CEREC Omnicam immediately after use as follows:
1.
Carefully wipe the mirror sleeve while it is on the camera so that no
dirt whatsoever can remain stuck and can harden on the surface of
the mirror sleeve. Use an absorbent cotton-based gauze for this
with water and neutral soap or a lint-free cloth that has been soaked
in one of the cleaning agents listed under "Cleaning agents and
disinfectants [→ 3]".
2.
Wipe down the mirror sleeve afterwards with the absorbent cotton
gauze dipped in drinking water.
3.
Next dry the mirror sleeve using a lint-free cloth.
2.6
Wipe disinfection for the camera and mirror
sleeve
NOTE
Do not spray the CEREC Omnicam or immerse it in cleaning agents
or disinfectants!
1.
Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve [→ 6]").
2.
Use a new absorbent cotton gauze which is moistened with one of
the agents listed in the "Cleaning agents and disinfectants [→ 3]"
section. First wipe the camera enclosure and then the mirror sleeve.
The camera enclosure and mirror sleeve are disinfected through
this step.
3.
Wipe down the mirror sleeve afterwards again with the absorbent
cotton gauze dipped in drinking water.
4.
Next dry the mirror sleeve using a lint-free cloth.