8 Maintenance
8.2 Calibrating CEREC Omnicam
Dentsply Sirona
Operating Instructions (valid for Canada)
70
66 37 487 D3652
D3652.201.01.21.02 03.2021
Measuring the position of the exit window
1.
Mount the bottom side of the calibration set to the tip of the camera.
2.
Click on the
"OK" button.
Ä
The calibration process is continued.
Ä
Once the calibration is complete, a message is displayed
indicating this.
3.
Confirm the message by clicking the
"OK" button.
Ä
The CEREC Omnicam is calibrated.
Error message during calibration
The software indicates if an error occurs during calibration. If the
calibration process resulted in errors, restart the process.
End calibration
ü
The software indicates that the calibration was completed
successfully.
➢ Click on the
"OK" button.
Ä
The CEREC Omnicam is calibrated.