4 Safety
4.2 Safety labels
Dentsply Sirona
Operating Instructions (valid for Canada)
14
66 37 487 D3652
D3652.201.01.21.02 03.2021
4.2
Safety labels
Plug connections of external PC interfaces
CAUTION
Additional devices connected to external interfaces must be tested
according to the relevant standards, e.g.:
EN 60601-1:2006 + Cor.:2010 + A1:2013,
IEC 60601-1 Edition 3.1:2012,
EN 61010-1:2010 based on IEC 61010-1:2010 + Cor.:2011.
They must be installed outside of the patient area (a radius of 1.5m
surrounding the patient).
CAUTION
Low voltages are applied to the coupling box for connecting external
interfaces.
➢ Do not touch the pins of the connectors.
NOTE
The externally connected cables must not be subjected to pulling
stress.
CAUTION
Trip/fall hazard
When installing the supply cable (the cable between the coupling box
and the camera storage cradle) there might be a risk of tripping.
➢ Lay the cable so that there is no risk of tripping.
➢ Attach the supply line so that it remains fixed at all times.
CAUTION
To maintain electrical safety, the PC must not be operated within the
patient area (a radius of 1.5m surrounding the patient).