Dentsply Sirona
Operating Instructions (valid for Canada)
8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
66 37 487 D3652
D3652.201.01.21.02 03.2021
63
English
10. Place the lid on the container and leave the mirror sleeve at least 12
minutes in the disinfectant. While exceeding the recommended
immersion time may not cause damage to the mirror sleeve, the
mirror sleeve should be removed as soon as possible after the
recommended time.
11. Remove the lid and the holder with the mirror sleeve from the
container.
12. Thoroughly rinse the mirror sleeve with tap water of potable water
quality that meets Federal Clean Water Standards for at least 30