SECTION 5: INSTALLATION AND SETTING
S
Setting temperature for heating needle.150
℃
、
180
℃
、
200
℃
、
220
℃
POWER
Extended press (2-3 seconds)
for switching on/off the device,
short press for starting/stopping the heating. Device will power off
automatically when heating more than 15 minutes or no operation
in 5 minutes.
Remarks
After temperature setting finished, please short press the POWER
button to start heating.
Trigger
Pushing the plugger to inject the
gutta-percha
LCD display
Set temperature, heating temperature
display and battery status display.
Symbol
:
-------Warning-up
-------Energy-saving function
Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
10
Power status
:
-------Low battery -------Full power
-----Cleaning status, constant temperature is 100
℃
,quit
cleaning function by press any button.
-----Energy-saving status, constant temperature is 120
℃
,
quilt this function when hold the inductive switch or
press button “S”(this mode can be shut off manually)
Remarks
When“
”changed to “E1”means heating malfunction ,“E2”means
warming-up malfunction. With the device off
,
extended press button
POWER until 2 beeps
,
energy saving function can be turned on or off. It
doesn’t show
for turning on the device when this function is off,
you can check if the
is showed on the status bar when turning on
the device.
Inductive switch
In the handshake position, the system is
equipped with inductive switch. The sense of the
switch should be timely and accurate to detect if
the doctor's hand is holding or loosening the
device in order to save the energy and raise the
temperature quickly.
Needle Nut
Fix the injection needle