Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
21
B. Backfilling
1. Put the gutta percha bar inside the heating chamber and then forwarding
the plunger to the front until no move can be felt. Then warm the gutta percha
bar to the setting value.
2. After the gutta percha has been warmed up and then insert the injection
needle of the gun as far as it will go into the canal space without binding.
3. Activate the trigger and inject the warmed gutta percha into the canal until
orifice by feeling the injected gutta percha push the tip out of the canal.
4. Compact the warmed gutta percha firmly by using the hand squeezing
device.
5. Repeat previously mentioned 3-4 procedures until the entire canal is filled
completely.
6. Confirm obturation with a radiograph.
SECTION 8: CLEANING & MAINTANANCE
A.
FreeFill Obturation Gun
It’s recommended that the gun should be cleaned after every use to prevent
built-up of excess gutta percha in the cylinder.
1. Adjust the unit into operating temperature. Press the plunger
release button
and pull the plunger out. Using a towel to wipe off gutta percha found on the
plunger.
2. Turn the unit off and unplug the power cord. Let the gun be cooled before
cleaning.
3. Insert the plunger and re-assemble the obturation gun.
B.
FreeFill Obturation Pen
Clean the tips regularly with the use of a towel with the cleaning solution
provided. Do not open the pen handpiece in any case.
If you want to clean the pen tip during operation (Heating state), please use
dried towel, do not use wet cleaning solution.
Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
22
SECTION 9. TROUBLE SHOOTING
TROUBLES CHECK
Can not be
recharged
Check the power indicator
Check that the adapter and charging base are
connected properly
Check that the battery and charging base are
connected properly
Check that the adapter and power socket are
connected properly
No reaction after
pressing button
PWR
Check that battery is no-power.
Please try to press button extendedly
Check that the installation of battery is correct
Check that the battery is in good order
Pen tip does not
heat.
Check that pen tips are in good order.
Check that activate heating operation
Check that pen tips are firmed properly
No GP coming out
of injection needle
Check that injection needle is blocked. If yes, clean the
injection needle or change the new one.
Check that the temperature comes up to the required
degree.
E1 for FreeFill Gun
Restart the device, if happen again, please contact the
local dealer or manufacturer
E2 for FreeFill Gun
Please check if the battery is sufficient in power. If the
electricity power is enough, please contact the local
dealer or manufacturer
E1 for FreeFill Pen
Heat tips may burn out, please replace the heat tips, if
it can not be solved, please contact the local dealer or
manufacturer
It’s not recommended to open or disassemble the control unit, pen and
gun in any case and manufacturer or local dealer does not have any
responsibility on problems or troubles caused by opening or