B. Backfilling 

1. Put the gutta percha bar inside the heating chamber and then forwarding 
the plunger to the front until no move can be felt. Then warm the gutta percha 
bar to the setting value. 
 
2. After the gutta percha has been warmed up and then insert the injection 
needle of the gun as far as it will go into the canal space without binding. 
3. Activate the trigger and inject the warmed gutta percha into the canal until 
orifice by feeling the injected gutta percha push the tip out of the canal. 
4. Compact the warmed gutta percha firmly by using the hand squeezing 
device. 
5. Repeat previously mentioned 3-4 procedures until the entire canal is filled 
completely. 
6. Confirm obturation with a radiograph. 
 

SECTION 8: CLEANING & MAINTANANCE     

 
A.

 

FreeFill Obturation Gun 
 

It’s recommended that the gun should be cleaned after every use to prevent 
built-up of excess gutta percha in the cylinder. 
1. Adjust the unit into operating temperature. Press the plunger

 

release button 

and pull the plunger out. Using a towel to wipe off gutta percha found on the 
plunger. 
2. Turn the unit off and unplug the power cord. Let the gun be cooled before 
cleaning. 
3. Insert the plunger and re-assemble the obturation gun. 
 

B.

 

FreeFill Obturation Pen 
 

Clean the tips regularly with the use of a towel with the cleaning solution 
provided. Do not open the pen handpiece in any case. 
If you want to clean the pen tip during operation (Heating state), please use 
dried towel, do not use wet cleaning solution. 

     

Rev. 02/01/18      VER SMS-DYGP-PLUS-FREEFILL20170828-EN

 

 

22 

 

SECTION 9. TROUBLE SHOOTING   
 

TROUBLES CHECK 

Can not be 
recharged

 

Check the power indicator 
Check that the adapter and charging base are 
connected properly 
Check that the battery and charging base are 
connected properly 
Check that the adapter and power socket are 
connected properly 

No reaction after 
pressing button 
PWR

 

Check that battery is no-power.   
Please try to press button extendedly     
Check that the installation of battery is correct   
Check that the battery is in good order

Pen tip does not 
heat. 

Check that pen tips are in good order.
Check that activate heating operation 
Check that pen tips are firmed properly

No GP coming out 
of injection needle

Check that injection needle is blocked. If yes, clean the 
injection needle or change the new one.   
Check that the temperature comes up to the required 
degree.  

E1 for FreeFill Gun

Restart the device, if happen again, please contact the 
local dealer or manufacturer   

E2 for FreeFill Gun

Please check if the battery is sufficient in power. If the 
electricity power is enough, please contact the local 
dealer or manufacturer 

E1 for FreeFill Pen

 

Heat tips may burn out, please replace the heat tips, if 
it can not be solved, please contact the local dealer or 
manufacturer 

 
It’s not recommended to open or disassemble the control unit, pen and 
gun in any case and manufacturer or local dealer does not have any 
responsibility on problems or troubles caused by opening or 

Содержание DY-GP Plus

Страница 1: ...FREEFILL20170828 EN USER S MANUAL DY GP Plus Cordless GP Obturation System The unit must be installed by a qualified engineer Only for user by dental professionals Read this operation manual carefully before installation or operation ...

Страница 2: ...need to notice the user in advance CONTACT INFORMATION The device is manufactured by DENJOY DENTAL CO LTD Address F4 Building A4 Lugu Medical Device Park No 229 Guyuan Road Changsha 410205 P R China Website www denjoy cn E mail denjoy denjoy cn Phone 86 731 85126543 Fax 86 731 88885317 Authorized European Representative Company name LANDLINK GMBH Address DORFSTRASSE 2 4 79312 EMMENDINGEN GERMANY P...

Страница 3: ... for injection needle z Faster power charging shorter charging time z Smart power saving mode offers longer operation z Light and more comfortable more ergonomic design z All parts which touched the patients can be autoclaved z Two in one system makes filling easier and more convenient z Digital display of temperature regulation it is more convenient z Special mode for changing injection needle an...

Страница 4: ... be used to cut off excessive gutta percha point SECTION 3 SYMBOL The following symbols may appear in this manual on the label or on it s accessories Some of the symbols represent standards and compliances associated with the device and its use Consult accompanying documents Caution Authorized Representative in the European Community Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 6 CE Mark con...

Страница 5: ... 18 VER SMS DYGP PLUS FREEFILL20170828 EN 8 Injection Device FreeFill Obturation Gun FreeFill Gun Injection device For backfilling the melted gutta percha Plugger Pushing gutta percha outside of the needle Status Description of the Plugger Gutta percha is pushed inside of the injection needle please start heating and backfilling when set temperature reached Gutta percha is used up please start the...

Страница 6: ...battery Full power Cleaning status constant temperature is 100 quit cleaning function by press any button Energy saving status constant temperature is 120 quilt this function when hold the inductive switch or press button S this mode can be shut off manually Remarks When changed to E1 means heating malfunction E2 means warming up malfunction With the device off extended press button POWER until 2 ...

Страница 7: ...s the red and green alternately to indicate that the battery is to be charged not charged Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 12 3 Place the injection device onto the charging base with the trigger inward match well the charging electrode the charging lamp is red and bright 4 The power is full when the charging lamp is green Remarks The charging base is not waterproof the liquid con...

Страница 8: ...g points to use this system efficiently and to avoid any mechanical trouble or dangerous condition Before using GP Obturation System the dentists need to make good root canal preparation with endo motor Requirement for good root canal preparation 1 All the root canal should be made as straight as possible 2 Root canals especially small collateral root canals should be cleaned as clean as possible ...

Страница 9: ...t s provides easy and fast filling but it s recommended to use the pen to fill the apical part at first to avoid any overfilling or under filling and the gun can be used for filling the remained part Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 16 of the canal 1 Select the appropriate injection needle and attach it to the gun Tighten firmly using the accessory tools provided Then put the bat...

Страница 10: ...on needle to avoid burning Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 18 B FreeFill Obturation Pen The pen is designed for vertical condensation or down packing with warmed gutta percha in the canal Besides it can be used to cut off excessive gutta percha point 1 Dry root canal Apply a thin layer of root canal sealer to the root canal wall and gutta percha points then insert the gutta perc...

Страница 11: ... trouble Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 20 SECTION 7 APPLICATION OF ROOT CANNAL FILLING First Phase Preparation 1 Gutta percha point Master cone Select the appropriate gutta percha point as a master cone which has same taper and size with prepared root canal Shorten 0 5 1 mm of the point than working length to avoid overfilling and to secure enough space to fill additional poin...

Страница 12: ...y case If you want to clean the pen tip during operation Heating state please use dried towel do not use wet cleaning solution Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 22 SECTION 9 TROUBLE SHOOTING TROUBLES CHECK Can not be recharged Check the power indicator Check that the adapter and charging base are connected properly Check that the battery and charging base are connected properly Ch...

Страница 13: ... within warranty period Do not modify and disassemble the device This instrument described below has been fully inspected and confronts to the current products specification The control part is guaranteed for its designated use against original defects in materials and workmanship for a period of 12 months from date of purchase The guarantee for spare parts is 6 months from date of purchase Produc...

Страница 14: ... environment specified below The customer or the user of the DY GP Plus should assure Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 26 that it is used in such an environment Immunity Test IEC 60601 Test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile...

Страница 15: ...diated RF IEC 61000 4 3 80MHz to 2 5GHz d 1 2 d 1 2 80MHz to 800MHz d 2 3 800MHz to 2 5GHz where p is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in ...

Страница 16: ...mitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 80MHz to 800MHz d 1 2 800MHz to 2 5GHz d 2 3 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3 8 7 3 100 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where...

Отзывы: