_______________________________________________
Operating (Set Point) Range:
0
°
C to 45 °C
Ambient Temperature:
10
°
C to 40 °C (non-condensing)
Stability / Repeatability:
+ 0.1 °C with constant load (even near ambient)
Cooling Capacity:
180 watts @ 20°C in a 20°C ambient
(See figure 1 for cooling curve)
Heating Capacity:
>240 watts
@ 20°C in a 20°C ambient
Noise (at 1 meter):
< 55 dBA
Coolant Type:
27% Propylene Glycol and Water recommended preventing
bacterial growth and minimizing noise and extending pump
life. (Note: propylene glycol is non-toxic). SSCS
Recommends Koolance LIQ-702CL-B coolant.
Ethylene Glycol and Water mixtures are also acceptable.
Note: Methanol, Ethanol, or Isopropyl Alcohol may not
be used as coolants, either by themselves or in water
mixtures; use of these coolants voids the chiller
warranty.
Process Fluid Fittings:
1/8” CPC quick disconnect with shut-off valves
Coolant Pump:
500ml/min brushless DC magnetically driven gear
(See Figure 2 for pump curve)
Coolant Outlet Pressure:
Nominal 10psi @ 500ml/min, Maximum 35 psig
Tank Volume:
75 ml
Wetted Materials:
Copper, polymers and stainless steel
Size (L x W x H):
7.5” x 5” x 7” (19 x 13 x 18 cm)
Weight:
9 lbs (4.1 kg)
Operating Voltage:
13.5VDC, 15 amps max
AC/DC Power Supply Output:
200 watt, 13.5 VDC; 100-240 VAC input
AC/DC Power Supply Input:
Universal 100-240 VAC, 50/60 Hz, 2.8 amps max
Communications:
Keypad or RS232 interface
Alarms:
Temperature, fluid level, system or component failure
Safety Certifications: (with power supply)
CE/ UL (TUV)
EN 61010-1: 3
rd
Edition
UL 61010-1: 3
rd
Edition
CAN/CSA 61010-1: 3
rd
Edition
RoHS
RoHS Compliant