Figure 4: MENU STRUCTURE:
N
OMENCLATURE
:
UP or Increase Value
Down or Decrease Value
↵
Press Enter Momentarily
Press & Hold Enter Key 3 Sec
SIMPLE
OPERATION
ADVANCED
OPERATION
(S
TATUS
M
ENU
)
press and hold enter key
(P
ARAMETER
I
NPUT MENU
)
TEMP:
XX.X°C
(current temp)
PASSWORD
XXXX
P
RESS
OR
(change set point)
↵
SETTEMP1
XX.X°C
SETTEMP1
XX.X°C
↵
↵
TEMP:
XX.X°C
(current temp)
SETTIME1
XXXX
(in minutes)
↵
SETTEMP2
XX.X°C
↵
SETTIME2
XXX
(in minutes)
↵
#
CYC
XXX
↵
ALRM
TMP
+/-X.X°C
↵
OFFSET
+/-X.X°C
↵
CHNG
PWD
Y/N
(change
password)
Allowable Value Ranges:
SETTEMP1
2
TO
45
°C
SETTIME1
0
TO
999 minutes
SETTEMP2
2
TO
45
°C
SETTIME2
0
TO
999
minutes
#
OF
CYC
0
TO
999
ALRM
TMP
1
TO
9
O
C
in 0.1°C increments
OFFSET
-5
°C
TO
5
°C
in 0.1°C increments
CHNG
PWD
Y
OR
N
NOTE: If continuous control at one set-point temperature is desired, set # OF CYCLES to zero.