6.
TECHNICAL DATA
Measurement range:
0 – 32 %
Error / Accuracy
± 2 %
Shaft diameter
25 mm
Test glass
40 x 13 mm
Size
175 x 40 mm
Weight
140 g.
Operation Manual for Hand-held Refractometer for
Brix measurement 28 to 62%
Product Layout
Specification:
Model:
RBX2862
Measuring range:
28 ~62%
Display Resolution:
0.2%
Temperature Compensation: Automatic from 10 to 30degree Celsius.
(ATC model only)
Calibration:
1.
Calibration should be made in an ambient environment of 20degree Celsius.
2.
Open the Light Plate & place few drops of calibration fluid ½ fill the prism.
Model
Calibration
fluid
Calibration
Point
RBX2862
29.6% Sucrose Solution
29.6%
Alternative standards may be used.
3.
Close the Light Plate. Ensure the water is spread across the entire surface of
the Prism. There must not be any bubbles or dry spots on the Prism, which is
cause by too little liquid use.
4.
Wait for about 30 seconds, this is to allow the temperature compensation to
take place.
5.
Hold the meter in the direction of light source and look into the Eyepiece.
6.
Adjust the Focus Dial until image is clear and sharp. Each person has his own
focusing adjustment.
7.
Remove the rubber cap on the Calibration knob, with the use of the provided
trimming tool and adjust it until the bottom line of the blue section is at 29.6%.
8.
Calibration Completed.
9.
Wipe dries the Light Plate and Prism with soft lint-free clothes before
proceeding with measurement.
Operation Manual for Hand-held Refractometer for
Brix measurement 28 to 62%
Product Layout
Specification:
Model:
RBX2862
Measuring range:
28 ~62%
Display Resolution:
0.2%
Temperature Compensation: Automatic from 10 to 30degree Celsius.
(ATC model only)
Calibration:
1.
Calibration should be made in an ambient environment of 20degree Celsius.
2.
Open the Light Plate & place few drops of calibration fluid 1.2 fill the prism.
Model
Calibration
fluid
Calibration
Point
RBX2862
29.6% Sucrose Solution
29.6%
Alternative stanards may be used.
3.
Close the Light Plate. Ensure the water is spread across the entire surface of
the Prism. There must not be any bubbles or dry spots on the Prism, which is
caused by too little liquid use.
4.
Wait for about 30 seconds, this is to allow the temperature compensation to
take place.
5.
Hold the meter in the direction of light source and look into the Eyepiece.
6.
Adjust the Focus Dial until image is clear and sharp. Each person has his own
focusing adjustment.
7.
Remove the rubber cap on the Calibration knob, with the use of the provided
trimming tool and adjust it until the bottom line of the blue section is at 29.6%.
8.
Calibration Completed
9.
Wipe dries the Light Plate and Prism with soft lint-free clothes before
proceeding with measurement.
Eyepiece
Calibration Knob
Focus Dial
Light Plate
Prism
Eyepiece
Calibration Knob
Focus Dial
Light Plate
Prism