6
1. Depressurize the purifier.
2. With indicator still installed, unscrew combination bleed
and check valve assembly from indicator housing.
3. Remove lower gasket and old indicator cartridge.
4. Replace upper gasket and check to insure that it is prop-
erly seated in housing.
5. Install new indicator cartridge in housing so that thinner
band of colored granules (indicator gel) is at the top.
6. Insert new lower gasket.
7. Screw the combination bleed and check valve assembly
back into indicator housing.
6.0 GENERAL INFORMATION
1. When the purifier is not in operation, all valves should be
closed, including the vessel drain cocks and the indicator
bleed valve. The convertor cartridge and indicator vessel
contain moisture sensitive materials.
NOTE:
When indicator bleed valve is closed and system
is not in operation, an integral check valve will prevent
atmospheric air from entering and contaminating car-
tridges.
2. Replacement convertor cartridge should not be exposed
to moisture or high humidity. Do not unseal vapor bag
until immediately before installation and operation.
3. Replacement of the convertor cartridge, which is con-
sumable, is much more frequent (hours of use) than the
filter cartridge (months of use).
4. Consult factory for questionable applications or when in
doubt about intended usage.
PARTS LIST
Description
Part
Number
Cartridge
Replacement No.
Filter
Convertor Case
Moisture Indicator
3154029
3154185
4003168
0713-2-1
3154050
3154165
Isolating Valve (2)
Drain Cock (Manual) (2)
Bleed & Check Valve Assembly
Spring
4009899
3154103
3154199
3154187
Bracket
O-ring 1-3/4 ID x 1-7/8 OD
Nut Hex 5/16-18
Sleeve Seal Assembly
O-ring 1-7/8 ID x 2 OD
3154034
3154404
3154135
3154411
3154405
Inlet
Orifice
Indicator
Housing
Upper Gasket
Indicator
Tube
Lower Gasket
Combination Bleed &
Check Assembly
INLET
ORIFICE
INDICATOR
HOUSING
UPPER
GASKET
INDICATOR
TUBE
LOWER GASKET
COMBINATION
BLEED & CHECK
VALVE ASSEMBLY
5.3 Indicator Kit P/N 3154165