DELTA
LTL2000S(Q) Retroreflectometer
35
Command set
Command
Parameter
Response
Meaning
TO
None
Sensor Off
Turn sensor off (power off)
VAn None
11.50
VBat Alarm: 11.00 Volt
VBat Alarm: 11.50 Volt
Set point for low bat warning
New set point for low bat warning
VB
None
VBat = 12.77 Volt
Measure actual voltage on lead-acid battery
VS
None
VBat Lamp Off: 12.78 Volt
VBat Rl Lamp On : 11.99 Volt
VBat Qd Lamp On : 12.09 Volt
Measured VBat with min. load
Measured VBat with Rl. lamp load
Measured VBat with Qd lamp load
<?>
None
DA [yyyy mm dd]
DPS x-stat
FV Firmware Version
ID ID
II Init [Y/N]
LC Log Clr
LE Log , sep
LS Log stat
LWn Log warn [T/F]
OT Off timer
PV PMTV
QEn Qd[T/F]
QCn Cal to n
REn Rl [T/F]
RCn Calc to n
RZ zero
M Measure
SNn Seq #
TI [hh:mm:ss]
TO Turn Off
VA Bat alarm
VB VBat
VS V-idle V-lamp
Help menu.
Query/Set Real Time Watch Date year/month/date
Query instrument status
Query Firm Ware info
Instrument S/N
Enable/Execute Instrument Initialisation
Clear Log Data
Send Log Data in comma seperated form
Query Data Log Status
Enable/Disable the Display of Log full warning [T/F]
Query/Set the Automatic Turn Off Timer
Read PMT Voltage
Enable/Disable Qd function
Calibrate Qd to n
Enable/disable Rl Function
Calibrate Rl to n
Zero adjust Rl measurement
Execute enabled measurements
Query/Set Sequence ID
Query/Set Time
Turn instrument Off
Query/Set Alarm for Low VBAT
Query Battery voltage
Query Voltage conditions