Delta Electronics Servo Press S-Series Installation and Maintenance Instructions
19
5.2
Load Cell Calibration
In order for the servo press to maintain precise quality, please perform load cell calibration if unusual
force levels appear, or perform load cell calibration on a regular basis (after the warranty period
expires). The load cell calibration tool is shown in the figure. Please follow the Operating Instructions
while performing calibration and qualification or contact the Company (Delta Electronics) or contract
service providers.
5.3
Periodic Inspection
Servo press maintenance and inspection shall be carried out based on the usage conditions; please
contact our company (Delta Electronics) or contract service providers for the maintenance and
inspection.