SM150T User Manual 1.0
Operation
14
Logger connections and configuration
GP1 Logger
Two SM150Ts can connect to each GP1.
Each soil moisture sensor is wired as a differential,
powered sensor.
Requirements
GP1 logger (with v1.48 firmware or later)
PC running DeltaLINK (version 3.6 or later)
SM150T with SMSC/lw-05 cable
Channel 1 and 2 wiring
SM150T wire
Colour GP1 terminal
Power 0V
brown
CH1
/2
(GND)
Power V+
white
CH1
/2
(PWR)
Signal HI
blue
CH1
/2
(+)
Signal LO
black
CH1
/2
(-)
Cable shield
green
CH1
/2
(GND)
Tempe
grey
Temp3 (IN)
Using the DeltaLINK2 logger software, configure channel 1 or 2 as soil
moisture sensor type
SM150T
channel 3 or 4 as
SM150T Temperature
.
Two more SM150T sensors can be added, to Temp3 and/or Temp4 channels
to measure soil moisture -if temperature readings are not required
:
Power 0V
brown
CH1
/2
(GND)
Power V+
white
CH1
/2
(PWR)
Signal HI
blue
CH1
/2
(+)
Signal LO
black
CH1
/2
(-)
Cable shield
green
CH1
/2
(GND)
See also
GP1 Quick Start Guide
v4 or later and the
DeltaLINK
on-line Help.
2
You need the PC logger software DeltaLINK version 3.6 or later. A free upgrade can
Software and Manuals DVD
.