23
The wires coming from the
conductivity probe
are welded directly on the DIN45326 connector as
reported in the following table:
Sensor
Direct connection to the connector DIN45326
Jumper
2-electrode
conductivity
probe
Internal view of the connector
Jumper between
pins 1 and 4
Jumper between
pins 2 and 5
4-electrode
conductivity
probe
Internal view of the connector
None
As an alternative, the
Pt100 and Pt1000 sensors
can be welded directly on the connector's pins
DIN45326, as reported in the following table:
Sensor
Direct connection to the connector DIN45326
Jumper
Pt100 4 wires
Pt100
Internal view of the connector
None
Pt1000 2 wires
Pt1000
Internal view of the connector
Jumper between
pins 6 and 7
Jumper between
pins 3 and 8