background image

-      - 

35

CONNECTION

 

TO

 

A

 

PC

 

 

HD2106.1 

Connection to the PC with the cable: 

 

HD2110CSNM

:

 

8-pole MiniDin connector on one end and 9-pole Sub D connector on the 
other end; 

 

C.206

: 8-pole MiniDin connector on one end and USB type A connector on the other end. 

With integrated RS232/USB converter (requires the installation of the USB drivers). 

 

HD2106.2 

Connection to the PC with the cable: 

 

CP23

:

 

Mini-USB type B connector on one end and USB type A connector on the other end; 

 

HD2110CSNM

:

 

8-pole MiniDin connector on one end and 9-pole Sub D connector on the 
other end; 

 

C.206

: 8-pole MiniDin connector on one end and USB type A connector on the other end. 

With integrated RS232/USB converter (requires the installation of the USB drivers). 

 
The instruments are supplied with the DeltaLog9 software that manages the connection, data 
transfer, graphic presentation, and printing operations of the captured or logged measurements. 

The DeltaLog9 software is complete with "On-line Help" (also in PDF format) describing its 
characteristics and functions. 

 

 

C

ONNECTION TO THE 

RS232C

 SERIAL PORT OF THE INSTRUMENT

 

1.

 

The measurement instrument must be switched off. 

2.

 

Using the Delta Ohm HD2110CSNM or C.206 cable, connect the measurement instrument to the 
first free RS232C (COM) or USB serial port of the PC.  

3.

 

Turn on the instrument and set the baud rate to 38400 (MENU >> ENTER until the Baud Rate 
parameter >> select 38400 using the arrows >> confirm with ENTER). The parameter remains in 
the memory until replacement of the batteries. 

4.

 

Start the DeltaLog9 application and press CONNECT. Wait for the connection to occur and fol-
low the indications on the screen. For a description of the DeltaLog9 application, please refer to 
its on-line Help. 

 

C

ONNECTION TO THE 

USB

 

2.0

 PORT OF THE INSTRUMENT 

-

 ONLY FOR 

HD2106.2 

The connection via the CP23 cable does not require the installation of USB drivers: when connecting 
the instrument to the PC, the Windows® operating system automatically recognizes the device as an 
HID device (Human Interface Device) and uses the drivers already included in the operating system. 
To check if the connection has been successfully completed, double-click on "

Device Manager

from the Control Panel. The following items should appear: 

"

Human Interface Device

"

 >> 

"

HID-compliant device

"

Human Interface Device

"

 >> 

"

USB Human Interface Device

When the USB cable is disconnected, the items disappear and reappear when it is connected again. 

Содержание HD2106.1

Страница 1: ...written in this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cann...

Страница 2: ...26 INSTRUMENT SIGNALS AND FAULTS 27 LOW BATTERY WARNING AND BATTERY REPLACEMENT 29 INSTRUMENT STORAGE 30 SERIAL INTERFACE AND USB 31 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 33 THE LOGGING...

Страница 3: ...ransferred from the instrument connected to a PC via the RS232C serial port or USB 2 0 port The storing interval printing and baud rate can be configured using the menu The HD2106 1 and HD2106 2 model...

Страница 4: ...4 Conductivity meter Thermometer HD2106 1...

Страница 5: ...etween the current value and the logged value when the key is pressed in the menu decreases the current value 9 SERIAL key starts and ends the data transfer to the serial communication port 10 MENU ke...

Страница 6: ...6 Conductivity meter Thermometer HD2106 2...

Страница 7: ...the serial communication port In the menu clears the data contained in the instrument s memory 10 LOG DUMP LOG key during normal operation starts and ends the saving of the data in the internal memor...

Страница 8: ...ally it sets the instrument ready for normal measurement MIN cm min s 2 3 m m k During turning on should no probes be connected the last manually set temperature appears in the secondary line The unit...

Страница 9: ...n of this MENU key below Attention the data captured using the Record function cannot be transferred to the PC CAL key It increases the current parameter when used in the menu when used in measurement...

Страница 10: ...ment of memorized data only HD2106 2 the message _LOG_DUMP_or_ERAS Transfer data or erase is scrolled in the comment line The center figure reports the number of free memory pages FREE Pressing SERIAL...

Страница 11: ...ero Pressing SERIAL only causes the print to occur when the measurement has stabilized HOLD symbol still Later it is possible to re peat the print at will but while the HOLD mode is on the sample iden...

Страница 12: ...conversion factor depends on the nature of the salts present in the solution in the field of water quality treatment and control where the main component is CaCO3 Calcium Carbonate a value of 0 5 is u...

Страница 13: ...and PC baud rates are the same If the USB connection is used this parameter value is automatically set please see the details on page 33 LOG DUMP LOG key only HD2106 2 In measurement mode this functio...

Страница 14: ...when using an external power supply the battery symbol does not appear Press SERIAL to end the continuous printing Before starting the printing with SERIAL set the baud rate To do so select the Baud R...

Страница 15: ...1413 S cm 25 C 0 1 Molar KCl solution 12880 S cm 25 C 1 Molar KCl solution 111800 S cm 25 C User calibration of the temperature sensors is not required The probes are detected during turn on and this...

Страница 16: ...low constant K probes are preferably used for low conductivity values the high constant ones for high values The indicative measurement range is reported in the following diagram S cm mS cm 0 1 0 1 0...

Страница 17: ...ation deletes the data of the previous calibration For example if the instrument is calibrated in 4 points and a new 1 point calibration is performed the instrument will be calibrated in 1 point only...

Страница 18: ...end the probe calibration press TDS ESC 9 Rinse the probe with water If you are then going to perform low conductivity measurements we recommend rinsing the probe using distilled or bidistilled water...

Страница 19: ...nt as it was before the calibration 13 Rinse the probe with water If you are then going to perform low conductivity measurements we recommend rinsing the probe using distilled or bidistilled water The...

Страница 20: ...0 124 1173 10 72 94 4 27 0 153 1467 13 37 115 7 17 0 126 1199 10 95 96 3 28 0 157 1494 13 62 117 7 18 0 128 1225 11 19 98 2 29 0 161 1521 13 87 119 8 19 0 130 1251 11 43 100 1 30 0 164 1548 14 12 121...

Страница 21: ...ensor is housed in the end part of the probe When measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to per...

Страница 22: ...ctivity cell are welded directly onto the connector s pins On the left there are the 1 4 points on which the Pt100 or Pt1000 sensor wires must be welded The JP1 JP4 jumpers are in the center of the ca...

Страница 23: ...ns 1 and 4 Jumper between pins 2 and 5 4 electrode conductivity probe Internal view of the connector None As an alternative the Pt100 and Pt1000 sensors can be welded directly on the connector s pins...

Страница 24: ...nnector Jumper between pins 1 and 4 Jumper between pins 2 and 5 4 ring conductivity probe Pt100 temperature sensor Pt100 Internal view of the connector None Ensure the welds are clean and perfect Once...

Страница 25: ...be held perpendicular against the surface Apply oil or heat conductive paste between the surface and the probe in or der to improve contact and reduce reading time Whatever you do do not use water or...

Страница 26: ...Check that the cell is not dirty that there are no air bubbles inside it Calibrate again using the appropriate standard Slow response or instability Check that the cell is not dirty that there are no...

Страница 27: ...ory full the instrument cannot store further data the memory space is full CAL blinking No calibration has yet been effected or the cell constant value has been modified in the menu please see the des...

Страница 28: ...OG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_DUMP_OR_ERAS transfer or erase data LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO SEC minutes...

Страница 29: ...lace the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2106 2 is logging and battery voltage falls below the minimum operating lev...

Страница 30: ...to direct sunlight The instrument may be exposed to a source of high temperature The instrument may be exposed to strong vibrations The instrument may be exposed to steam salt or any corrosive gas Ma...

Страница 31: ...m Cable Instrument port PC port Installation of USB drivers HD2110CSNM RS232 MiniDin RS232 9 pole SubD No C 206 RS232 MiniDin USB Yes CP23 USB Mini USB USB No The instrument standard serial transmissi...

Страница 32: ...the temperature n 0 C n 1 F LUBn Sets the unit of measurement for the conductivity n 0 S cm n 1 n 2 TDS n 3 NaCl RA Reading of LOG PRINT interval set RP 600 Battery level Resolut 0 01V RUA U C Temper...

Страница 33: ...upon pressing the LOG key the instrument logs the first data and turns off 15 seconds before the next logging instant it turns on again to capture the new sample and then turns off The data stored in...

Страница 34: ...a standard serial communication program e g Hyperterminal only through the RS232 serial port of the instrument using the cable HD2110CSNM or the cable C 206 Connect the HD40 1 printer using cable HD21...

Страница 35: ...rument must be switched off 2 Using the Delta Ohm HD2110CSNM or C 206 cable connect the measurement instrument to the first free RS232C COM or USB serial port of the PC 3 Turn on the instrument and se...

Страница 36: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Страница 37: ...2Vdc 1A Security of memorized data Unlimited independent of battery charge conditions Time Date and time Schedule in real time Accuracy 1min month max drift Measured values storing model HD2106 2 Type...

Страница 38: ...0 999 cm 1 cm 1 00k 19 99k cm 0 01k cm 20 0k 99 9k cm 0 1k cm 100k 999k cm 1k cm 1 10M cm 1M cm Measurement range Kcell 10 Resolution 0 5 5 0 cm 0 1 cm Accuracy resistivity 0 5 1digit Measurement of t...

Страница 39: ...conversion factor 0 4 0 8 Cell constant K cm 1 0 01 0 1 0 7 1 0 and 10 0 Standard solutions automatically detected 25 C 147 S cm 1413 S cm 12880 S cm 111800 S cm TECHNICAL DATA OF PROBES WITH MODULE I...

Страница 40: ...C 4 electrode cell Pocan Platinum SPT401 001 K 0 01 0 04 S cm 20 S cm 0 120 C 2 electrode cell AISI 316 steel SPT01G K 0 1 0 1 S cm 500 S cm 0 80 C 2 electrode cell Glass Platinum SPT1G K 1 10 S cm 1...

Страница 41: ...h in built temperature sensor RCT The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter HD22 2 Laboratory electrode holder composed of basis plate with incorporated magnetic stirrer st...

Страница 42: ...tor length 2 metres TP47 1000 O Immersion probe sensor Pt1000 Probe s stem 3mm length 230mm 2 wire connection cable with connector length 2 metres TP87 100 O Immersion probe sensor Pt100 direct 4 wire...

Страница 43: ...ali e nazionali garantita da una catena di riferibilit che ha origine dalla taratura dei campioni di prima linea dei laboratori accreditati di Delta OHM presso l Istituto Primario Nazionale di Ricerca...

Страница 44: ...2002 96 EC the European users of electrical and elec tronic equipment can return it to the dealer or manufacturer upon purchase of a new one The il legal disposal of electrical and electronic equipmen...

Отзывы: