18 / 53
4.2
P
ROGRAMMING WITH THE
D
ISPLAY
M
ODULE
The DMSP400 and DMSP500 should be adjusted to the process by programming the parameters. The Display Module can be used to display the parameters
during programming and measurement values during measurement.
The Display Module supports two separately accessible programming modes representing 2-layers of programming complexity, depending on user choice.
PLUG-IN DISPLAY MODULE
QUICKSET
FULL
PARAMETER
ACCESS
QUICKSET (5.2.4)
Recommended as a simple and fast way to set up the DMSP400/500 by 6 basic parameters
for the following basic settings, marked by abbreviations easy to remember
•
Engineering unit for the display (Metric or US)
•
Maximum measuring distance (H)
•
Assignment of min level to 4 mA
•
Assignment of max level to 20 mA
•
Error indication by the current output
•
Damping time
Full Parameter Access (5.2.5)
All features of the DMSP400/500 such as:
•
Measurement configuration
•
Outputs
•
Measurement optimisation
•
11 pre-programmed tank shapes for volume calculation
•
21 pre-programmed formula for flow metering
•
32-point linearisation