Expert
Logger
4.10.1.7
Measuring bridge sensor type
The example means: At a signal of 2.01523 mV/V, a force of 1000 N acts
on the sensor.
Scaling
Enter the
nominal load
and resulting signal of your sensor (
sensitivity
).
The output signal at nominal load is also referred to as gradient and is
noted in the calibration record of the sensor. For a linear characteristic,
measurement values are then converted into the unit of measurement
(scaling) from the sensor. For non-linear characteristics use
Sensor com-
pensation
"Sensor compensation tab" on page 50
tab. The spe-
cified range (in the example, ±1000 N) is used as the default for the
104
Delphin Technology AG
Januar 2018
Expert
Logger
Manual