YDT-35 Operating and Servicing Manual
44
HL066(EN) ISSUE 1, 05/2015
5 Troubleshooting
5.1 Software Error Codes
Software
Error Code
Error Description
Possible cause/Action
E000
Error present on unit.
Reboot the system if no other fault codes
are listed on the display.
E003
Protective guard cover open
Close the cover and restart test.
Faulty safety switches. Contact customer
support.
E100
Firmware/Database fault present
Contact customer support.
E101
PCB Over Temperature error
Allow PCB to cool down before continuing
test.
E105
Error in Database file
Contact customer support.
E110
Incorrect Injector profile selected
Refer to section 4.2.6 and select the
correct profile for the injector installed.
E200
High Voltage cannot be regulated
Refer to section 4.2.6 and select the
correct profile for the injector installed.
If problem persists, contact customer
support.
E300
Injector pulse error
Reboot the system by powering unit
off/on.
E400
Maximum pressure reached
Reduce pressure within limits and repeat
test.
E900
System Error
Reboot the system by powering unit
off/on.
If problem persists, contact customer
support.
E901
Error regulating voltage
Reboot the system by powering unit
off/on.
If problem persists, contact customer
support.
E902
E903
Injector High Voltage Error
Reboot the system by powering unit
off/on.
If problem persists, contact customer
support.
E990
Software Error
Reboot the system by powering unit
off/on.
If problem persists, contact customer
support.