16
o
The first line indicates the specific MC job.
o
The second line indicates the number of readings present in the specific job.
o
The third line shows the 95% Confidence Interval for the specific job. This is the value (+/-)
away from the calculated average value within which 95% of the stored readings are
guaranteed to exist.
o
The fourth line shows the value of the lowest MC reading stored in the specific job.
o
The fifth line shows the value of the highest MC reading stored in the specific job.
There are no control entries present in this screen, such as “MORE”, “CANCEL” or “DONE”. The only
available option is to return to the storage menu screen, by pressing any key.
o
Stored Readings
: This menu option selects the measurement review feature of the meter,
allowing the user access to measurements stored in memory. Only readings belonging to the
selected job will be retrieved.
NOTE:
Stored readings can only be reviewed in ascending sequence (ex: reading 1, 2, 3, etc.).
The meter assists moving to a set of readings later in the sequence by allowing the user to
specify a start point for this ascending sequence, such that the readings of interest will be
displayed first. The reading review process may be abandoned at any time.
o
Show MC Readings
: This feature selects stored MC readings for review. Choosing this feature
allows the user access to a job overview screen. This screen always begins by displaying the
current job, whether it contains any readings
or not.
The Left and Right buttons can be used to navigate sequentially through all available job
numbers (1-98) while the “JOB:” entry is selected. Jobs that are undefined or empty will
display “no readings”. Use the Up or Down buttons to change the line selection on the screen.
An empty or undefined job will only have one other selectable option: CANCEL. Choosing this
option and pressing the Select (center) button will bring the user back to the Storage Menu.
A selected job containing readings offers two other selectable options: “VIEW SET” and “DONE”.