background image

 

11 

 

ALARM  SETTING: 

 

Click SELECT on the ALARM  to switch the ALARM  on and off. If  “on”, a Value%  is shown. Use the 

left/right switches to lower or increase, and select DONE  to return to the MC READ  screen 

 

The “ALARM”  feature in the meter  provides the user with a set -point: readings  above a specified value 

(considered “WET” or unacceptable) generate an audible  alarm. 

The first reading  breaking  the alarm  set-point will  be signaled  by four short beeps in quick sequence. 

Subsequent readings  that remain  above the set -point will be signaled  by a single  short beep.  A 

reading  below the threshold  will cause the alarm  signal  to reset, such that any following  reading  above 

the threshold  will be once again  signaled  by four beeps. 

Once ALARM is chosen, pressing  the SELECT key toggles the state of the alarm  feature (ON-OFF). The 

ON state is marked  by the current alarm  threshold  value, while  the OFF state is indicated with an “OFF” 

statement. When the alarm  is ON, pressing  the LEFT or RIGHT  keys will change the threshold  value 

(down and up, respectively). Threshold values range from  0.05%  MC to 59.5%  MC (default is 8%). 

Select “DONE”  to return to the Create Job menu. 

Selecting “DONE”  in the Create Job menu  returns the user to the MC READ  screen. The meter  is now 

configured with the new job and all parameters  defined in the “Create Job” screen. Selecting “CANCEL” 

in the Create Job menu  returns the user to the Job Select screen; where it has the opportunity to re -

start the job select/create process. Pressing CANCEL  in the Job Select screen brings  the user back to 

MC READ.  The meter  maintains  the configuration  it had before the JOB  option was selected. 

 

 

 

 

Содержание RDM3-P

Страница 1: ...RDM3 P Paper Moisture Meter MANUAL Version 1 0 Rev 02 13 DELMHORST EUROPE TITANIUMLAAN 100 NL 5221 CK s HERTOGENBOSCH THE NETHERLANDS info Delmhorst eu www MoistureMetersDelmhorst com 31 0 73 6395080...

Страница 2: ...ING PAPER PAPER CORES AND CORRUGATED PRODUCTS 12 USING THE 0 100 RELATIVE SCALE 12 TESTING BALED SCRAP PAPER 13 APENDING DATA TO A JOB 14 STORAGE 14 STORAGE MENU 15 VIEW STATISTICS 15 STORED READINGS...

Страница 3: ...Three scales for o PAPER 5 3 23 o RELATIVE 0 100 scale and o BALED PAPER 8 40 NOTE Displayed readings outside of this nominal range should be taken as a no reading They are included in statistical ca...

Страница 4: ...nderside of the case at the bottom of the handle i Open the battery compartment by sliding the lid back while pressing on the release indent ii Ensure correct polarity and push the battery in flush wi...

Страница 5: ...efuse to stay on at power up This gradual warning system is intended to provide the user advanced battery status notice and give ample time for replacement before operational limitations occur NOTE St...

Страница 6: ...action changes the selected parameter s value for DONE and CANCEL parameters without assigned values the LEFT and RIGHT keys perform no function All parameter values are incremented using the RIGHT ke...

Страница 7: ...the battery is removed Meter Use This chapter describes each meter menu in detail as well as the use of each meter feature MAIN MENU The RDM 3P Main Menu is the default screen at power up It provides...

Страница 8: ...o a specific job More than one job can be defined with the same scale RDM 3P automatically configures MC READ Mode with the following default set of parameters when this feature is first accessed o De...

Страница 9: ...MODE option and pressing the SELECT key will change the measurement mode o Triggered mode causes a single reading to be taken every time the SELECT button is pressed Each reading is stored in memory T...

Страница 10: ...EAD Mode menu This is the same as selecting CANCEL Select JOB and use the LEFT or RIGHT keys to change the job number to the one desired When finished select DONE To abandon select CANCEL If the selec...

Страница 11: ...s Once ALARM is chosen pressing the SELECT key toggles the state of the alarm feature ON OFF The ON state is marked by the current alarm threshold value while the OFF state is indicated with an OFF st...

Страница 12: ...s not entirely embedded in the thickness of the sample the readings tend to indicate a lower than actual MC This can be overcome by testing more than one sample in stacks USING THE 0 100 RELATIVE SCAL...

Страница 13: ...or reject the material Since checking the moisture condition of bales is performed when buying and selling the specific value of the meter readings remains an element to be agreed upon between buyer a...

Страница 14: ...xisting jobs or create new ones at random There are no restrictions regarding measurement storage based on job number For example the user is free to take and store readings pertaining to job 3 then j...

Страница 15: ...cific job o The third line shows the Average MC value for all stored readings in the specific job o The fourth line shows the Standard Deviation for the specific job o The fifth line shows the Coeffic...

Страница 16: ...an only be reviewed in ascending sequence ex reading 1 2 3 etc The meter assists moving to a set of readings later in the sequence by allowing the user to specify a start point for this ascending sequ...

Страница 17: ...ately pressing the Left button causes the meter to subtract 10 from the current left side value o DONE Selecting this option signals to the meter that the user has completed the reading recall setup p...

Страница 18: ...to 30 cm 0 to 12 Operation 1 Begin by starting the Delmhorst Universal Application Program double click the application icon or chose the program from the Start Menu 2 Once the RDM3 and Base Unit are...

Страница 19: ...his is not the case check power connections o No connection to the PC The Base Unit must be connected to the PC through a serial cable Check connection b KBD This message occurs when the user presses...

Страница 20: ...our name address daytime phone and fax numbers or e mail address If you believe the meter is under warranty please provide the original sales slip or invoice Ship via UPS Express Mail Priority Mail or...

Страница 21: ...ctions in the Care of Your Meter section Under no circumstances shall DELMHORST be liable for any incidental indirect special or consequential damages of any type whatsoever including but not limited...

Отзывы: