ALARM SETTING:
Click SELECT on the ALARM to switch the ALARM on and off. If “on”, a Value% is shown. Use the
left/right switches to lower or increase, and select DONE to return to the MC READ screen
The “ALARM” feature in the meter provides the user with a set -point: readings above a specified value
(considered “WET” or unacceptable) generate an audible alarm.
The first reading breaking the alarm set-point will be signaled by four short beeps in quick sequence.
Subsequent readings that remain above the set -point will be signaled by a single short beep. A
reading below the threshold will cause the alarm signal to reset, such that any following reading above
the threshold will be once again signaled by four beeps.
Once ALARM is chosen, pressing the SELECT key toggles the state of the alarm feature (ON-OFF). The
ON state is marked by the current alarm threshold value, while the OFF state is indicated with an “OFF”
statement. When the alarm is ON, pressing the LEFT or RIGHT keys will change the threshold value
(down and up, respectively). Threshold values range from 0.05% MC to 59.5% MC (default is 8%).
Select “DONE” to return to the Create Job menu.
Selecting “DONE” in the Create Job menu returns the user to the MC READ screen. The meter is now
configured with the new job and all parameters defined in the “Create Job” screen. Selecting “CANCEL”
in the Create Job menu returns the user to the Job Select screen; where it has the opportunity to re -
start the job select/create process. Pressing CANCEL in the Job Select screen brings the user back to
MC READ. The meter maintains the configuration it had before the JOB option was selected.