background image

 

Copyright   2014, Delmhorst Instrument Co.  

Page 9 of 17 

 
 

TAKING A READING 

– PRACTICAL APPLICATIONS 

 

The following application notes are intended for use with the meter in PIN 
MODE except where otherwise indicated.  

  

TESTING WOOD: 

 

The contact pins on the 2-

E electrode are best for materials up to approximately 1” in thickness. 

On materials over 1”, we recommend using a probe such as the 26-ES ram-type electrode. Mount 
the 26-ES directly to the external connector. 

To take a reading

, align the contact pins parallel to 

the grain and push them to their full penetration into the wood, if possible. Insulated pins read 
only at the tip and can be driven to the desired depth.  

 

PAINT FAILURE AND MOISTURE 

 

Moisture is by far the most frequent cause of paint failure. The key to preventing paint failure is to 
insure that moisture is not absorbed through the wood to the back of the paint film. So, in order to 
insure quality paint jobs, wood must remain dry after the application of paint. 
 
Outdoor wood can be safely painted without danger of peeling if the %MC is 15% or less. In drier 
climates, the maximum reading should be 10% to 11%. Indoor wood should be between 7% and 
8% prior to painting.  
The following conditions may cause high moisture content in wood: 

 

  Leaky gutters and down spouts 
  Leaky pipes or condensation on cold water lines in attic or hollow walls 
  Faulty flashing around windows, doors and where porch and dormer roofs meet sidings 
  End-grain wood that is not sealed with paint at all joints around windows, corners, and butt 

joints 

  Porch columns that do not have good drainage and ventilation where they rest on porch 

floors 

  Siding or any other wood that is in contact with the ground may absorb moisture 
  Siding and shingles without sufficient lap so that water is forced up through cracks by wind 

pressure 

  Ice dams 
  Condensation of vapor within hollow walls 

 

EIFS (Exterior Insulation & Finish Systems)  

 

Moisture intrusion problems in EIFS (also known as synthetic stucco) stem from leaking window 
frames, improper use of or lack of sealant, and faulty installation of flashing. 
 
If you suspect a problem take a visual inspection. Look for gaps around windows, doors, air 
conditioning units, light fixtures, hose bibs, dryer vents and other areas of potential penetration. 
Also look for visible signs of water damage. If you feel a problem exists, use the 

21-E electrode

This electrode uses the 

608 (4") insulated pins or 608/001 (7”) insulated pins. 

 

Procedure: 

 

 

Drill two 1/4" holes

 

about ¾” apart at an upward 45° angle.  

 

 

Push the 21-E Electrode into the holes

 through the polystyrene and into the substrate and 

read the moisture content on the meter scale. When used on materials other than wood, the 
meter’s 

PIN

 mode will give relative readings only, not %MC. The numbers displayed 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание QuickNav 12545

Страница 1: ...Nav MOISTURE METER Owner s Manual Version 2 01 For Meters starting with Serial 12545 Delmhorst Instrument Co Rev 4 May 2014 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 2: ...O TAKE SCAN READINGS 7 TO TAKE TEMP RH READINGS 7 TO SET THE ALARM IN WOOD DRYWALL SCAN MODE 8 MAIN MENU OPTIONS 8 TAKING A READING PRACTICAL APPLICATIONS 9 TESTING WOOD 9 PAINT FAILURE AND MOISTURE 9 EIFS 9 10 TESTING CONCRETE 10 11 TESTING INSULATION 11 TESTING DRYWALL 12 CARE OF YOUR METER 13 SERVICE FOR YOUR METER 14 WARRANTY 15 SPECIES CORRECTION TABLE 16 17 www GlobalTestSupply com Find Qual...

Страница 3: ...r 10 90 Measures temperature over the range of 40 F 255 F with accuracy of 1 8 F over 4 F to 158F range of 40 C 124 C with accuracy of 1 C over 20 to 70 C Calculates Dew Point over the range of 40 F 176 F 40 C 80 C Calculates GPP Grains Per Pound when in Fahrenheit mode over the range of 0 1 3820 GPP Calculates GPK Grains Per Kilogram when in Celsius mode over the range of 0 01 545 GPK Calculates ...

Страница 4: ...he lid back while pressing on the release indent 2 Ensure correct polarity and push the battery in flush with the bottom board until the connectors snap together on both sides 3 Replace the battery compartment lid LOW BATTERY The meter features a battery status monitor designed to warn the user as well as protect measurement accuracy from impending battery failure conditions The battery warning is...

Страница 5: ...he keys to select the desired function mode and then press SELECT to activate the function Keypad Layout METER USE TO TAKE WOOD READINGS 1 From the Main Menu use the key to highlight WOOD Then use the SELECT key to enter the WOOD screen 2 The meter will display the WOOD screen as shown below 3 Push the contact pins into material to be tested Any Delmhorst electrode may also be used in WOOD mode by...

Страница 6: ... SELECT key to enter the pin mode 2 The meter will display the DRYWALL screen as shown below 3 Push the contact pins into the drywall Any Delmhorst electrode may also be used in DRYWALL mode by simply attaching it to the connector on top of the meter 4 The unit will read MC on a drywall scale between 1 and 6 5 To hold a reading on screen press the HOLD SELECT KEY Press the HOLD SELECT KEY again to...

Страница 7: ...into the top of the meter making sure to align the pin out on the DIN connector 1 From the Main Menu use the key to highlight RH T Then press the SELECT key to get into the RH T mode main menu The meter will display the main RH T mode main menu as shown below 2 The meter will begin taking readings Note You may change the temperature mode between F and C by highlighting the F C in the lower left ha...

Страница 8: ... alarm will sound if that value is reached Note Default alarm settings are WOOD 15 DRYWALL 1 SCAN 300 MAIN MENU OPTIONS METRIC IMPERIAL This option will toggle between Imperial units of measure and Metric units of measure when the meter is in RH T Mode Pressing SELECT key on this option when displayed as METRIC will change the units used to metric Pressing the SELECT key on this option when displa...

Страница 9: ...in attic or hollow walls Faulty flashing around windows doors and where porch and dormer roofs meet sidings End grain wood that is not sealed with paint at all joints around windows corners and butt joints Porch columns that do not have good drainage and ventilation where they rest on porch floors Siding or any other wood that is in contact with the ground may absorb moisture Siding and shingles w...

Страница 10: ...nsert the protective yellow cap into the sleeve and let the hole acclimate for 72h 5 Remove the cap and immediately insert the sensor to full hole depth Wait a minimum of 1h and then take RH readings by connecting one end of the RH T C1 cable to the top of the meter and the other end to the exposed connector on the sensor housing Check to ensure the meter s displayed RH reading is no long changing...

Страница 11: ...test Drill two 1 4 holes 3 4 apart and 1 2 to 2 deep Drive the masonry nails into the bottom of the holes and make the tests as described above Nails must not touch sides of drilled holes If the meter still indicates a dry condition the floor is ready for covering Tests should be made at several points especially when the slab is thick 4 or more and air circulation is poor Make tests only in newly...

Страница 12: ...two seconds USING SCAN MODE You can also take a reference reading on drywall using the meter s SCAN mode To take a reading firmly press the back of the meter onto the material to be tested while in SCAN mode This will display a relative reading that ranges between 0 and 300 Remember to establish a baseline reading by taking a reading in an area you know to be unaffected then take a reading in the ...

Страница 13: ... affected Under these conditions allow the meter to acclimate to the temperature conditions in which it will be used for minimum of 1 2 hours or as long as possible Change contact pins as needed Keep pin retainers hand tightened Clean the meter and contact pins with any biodegradable cleaner Use the cleaner sparingly and on external parts only Keep cleaner out of the external connector Remove the ...

Страница 14: ...r should stay in calibration follow the instructions in the Care of Your Meter section UNDER NO CIRCUMSTANCES SHALL DELMHORST BE LIABLE FOR ANY INCIDENTAL INDIRECT SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY TYPE WHATSOEVER INCLUDING BUT NOT LIMITED TO LOST PROFITS OR DOWNTIME ARISING OUT OF OR RELATED IN ANY RESPECT TO ITS METERS OR ELECTRODES AND NO OTHER WARRANTY WRITTEN ORAL OR IMPLIED APPLIES DEL...

Страница 15: ...19 20 5 22 5 24 HEMLOCK WESTERN 7 8 9 10 5 13 15 17 19 20 5 22 23 5 HACKBERRY 7 8 5 9 9 5 12 13 15 17 18 5 20 22 HICKORY 8 8 5 9 10 11 12 5 14 15 5 17 19 20 5 KERUING 8 9 10 11 13 15 17 20 22 24 27 LARCH 7 5 9 10 11 13 15 17 19 21 23 25 5 MAGNOLIA 7 5 9 10 11 5 14 16 17 5 19 21 22 5 24 5 MAHOGANY AFRICAN 8 9 5 10 5 12 15 17 19 5 22 24 26 28 ALSO KHAYA MAHOGANY HOND 7 8 9 10 5 12 5 14 5 16 18 19 5 ...

Страница 16: ...5 18 20 5 23 25 28 30 SPF COFI 8 9 10 11 13 15 17 19 21 23 25 TEAK 7 8 8 5 9 11 12 14 15 17 18 5 20 VIROLA 6 5 7 8 9 11 12 5 14 16 18 18 5 20 5 WALNUT BLACK 7 5 8 5 9 5 10 5 12 5 14 5 16 18 20 22 23 5 The species correction values shown in this chart have been rounded for easy reference Meter readings taken with 26 E 2 pin electrode Do not apply 2 pin correction SPF correction based on 2 pin 26 E ...

Отзывы: