AUSTROMAT
®
624 / 644
6.
Operation
Operating Instructions (Version 01/2012)
113
6.5.4. Programming
and program steps
Drying the firing object
The insulation table is still very hot immediately after firing. The temperature near the firing
object may thus be too high for drying. Wait a few minutes until the insulation table has cooled
down before placing the new firing object on the insulation table for drying.
Standby temperature
Do not worry if the firing furnace sets a stand-by temperature in the firing chamber that is lower
than the preheating temperature. If a drying time has been programmed, the furnace
automatically sets a temperature in the firing chamber that is necessary for the drying.
Cooling and relaxing
In the program steps "Cooling" and "Relaxing" (easy programs), the respective temperature is
approached as fast as possible. The firing chamber is opened in order to move the firing goods
out of the hot firing chamber. Since the firing chamber then cools, the fired materials are
subjected to less radiated heat than corresponds to the cooling temperature. To offset this
effect, the lift moves up slightly closer to the hot firing chamber. When the firing chamber has
reached cooling temperature, the lift should be completely at the top, and the firing chamber
should be closed again. This ensures that the fired materials are always approximately
subjected to cooling temperature during the cooling time.
6.5.5. Firing
programs
Firing tray
Every type of firing tray has an effect on the temperature of the firing object due to its color and
composition. Using a standardized firing tray can thus ensure constant firing results.
Firing parameters and test firings
Care in creating firing programs and editing firing parameters has a significant effect on the
quality of firing results. The recommended firing parameters are thus to be seen only as
guidelines. Perform test firings to identify your individual firing parameters based upon these
guidelines.
Содержание AUSTROMAT 624
Страница 1: ...Operating Instructions AUSTROMAT 624 644 Version 01 2012 ...
Страница 8: ...Table Of Contents AUSTROMAT 624 644 8 Operating Instructions Version 01 2012 ...
Страница 145: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 145 9 4 1 Kerform safety data sheet ...
Страница 146: ...9 Appendix AUSTROMAT 624 644 146 Operating Instructions Version 01 2012 ...
Страница 147: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 147 ...
Страница 148: ...9 Appendix AUSTROMAT 624 644 148 Operating Instructions Version 01 2012 ...
Страница 149: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 149 ...
Страница 150: ...9 Appendix AUSTROMAT 624 644 150 Operating Instructions Version 01 2012 ...
Страница 151: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 151 ...
Страница 152: ...9 Appendix AUSTROMAT 624 644 152 Operating Instructions Version 01 2012 ...
Страница 155: ...AUSTROMAT 624 644 10 Index Operating Instructions Version 01 2012 155 ...