TRIG?
Returns the trigger type (Internal, External, or Single Shot). Example:
TRIG?
INTERNAL
TRIG:INTERNAL, TRIG:EXTERNAL, or TRIG:SINGLESHOT
Sets the trigger type. The trigger types are Internal, External, or Single Shot.
Example:
TRIG:INTERNAL
response is
TRIG:INTERNAL
TRIG:EXTERNAL
response is
TRIG:EXTERNAL
TRIG:SINGLESHOT
response is
TRIG:SINGLESHOT
TRIG:EXT:TERM?
Returns the external trigger termination value. Example:
TRIG:EXT:TERM?
50 Ohms
TRIG:EXT:TERM n
Sets the external trigger termination of 50 ohms or 10k ohms, use n value of 50
or 10. Example:
TRIG:EXT:TERM 10
TRIG:EXT:TERM 10k
TRIG:INT:FREQ?
Returns the trigger frequency in hertz. Example:
TRIG:INT:FREQ?
1000000 Hz
TRIG:INT:FREQ n
Sets the trigger frequency in hertz. The value of n must be between 5 and
1000000.
Example:
TRIG:INT:FREQ 24000
TRIG:INT:FREQ 24000
TRIG:INT:MAIN:DUTYCYCLE?
Returns the duty cycle of the pulse in percent. Example:
TRIG:INT:MAIN:DUTYCYCLE?
28.10 %
Frequency
Resolution
5 Hz to 995 Hz
5 Hz
1.0 kHz to 49.9 kHz
100 Hz
50 kHz to 1.000 MHz
1000 Hz