CLEANING / MAINTENANCE
WARNING
•
•
•
•
DO NOT
allow water to enter any components, or Fill Connector.
DO NOT
immerse the Reservoir in any kind of liquid.
DO NOT
use any strong solvent or abrasive cleaners.
DO NOT
use alcohol, solvents, polishes, or any oily substance on Oxygen
Equipment.
1. Clean exterior surfaces with a cloth dampened with mild detergent and water.
2. Wipe dry with a clean cloth prior to use.
3. Store Reservoir in a clean area free from grease, oil, and other sources of
contamination.
4. Check and empty the Condensation Collection Container daily.
5. Clean the Condensation Collection Container with mild detergent and water, as
necessary.
RETURNS
Returned products require a Returned Goods Authorization (RGA) Number, contact
DEHAS
.
All Reservoirs must be depleted of liquid Oxygen prior to shipping
and must
be packaged in sealed containers to prevent damage.
DEHAS
, will not be responsible
for goods damaged in transit. Refer to
DEHAS
Return Policy available on the
DISPOSAL INSTRUCTIONS
Dispose of the Reservoir in accordance with the local regulations.
Please Recycle
TROUBLESHOOTING
PROBLEM
Portable will not fill
If the Reservoir fails to function, consult the Troubleshooting Guide below.
If problem cannot be solved, consult your liquid Oxygen Supplier.
PROBABLE CAUSE
REMEDY
1. Not connecting the Fill 1. make sure the Fill Connectors
Connectors completely are fully engaged
2. Contact liquid Oxygen 2. Reservoir empty
Supplier to refill Reservoir
Unable to disconnect
Fill connectors frozen Allow time for the Fill
the Portable from
from moisture on the Fill Connectors to thaw
Reservoir
Connectors
Bottom light flashes
low Battery Call your liquid Oxygen
Rapidly
Supplier for battery
replacement
No light(s) will show
Dead battery Call your liquid Oxygen
on the Liquid Level
Supplier for battery
Indicator
replacement
water leaking from
Condensation Container Empty the Condensation
Reservoir
is full Container
MINOR
LEAK
of Liquid
Frost in the Fill Connector Disconnect and reconnect
Oxygen
(Ref page 7) Portable
MAJOR
LEAK
of Liquid
Equipment Failure Call your liquid Oxygen
Oxygen
(Ref page 7) Supplier
8