DeguDent SOLARIS Service Manual
11 Troubleshooting
There are two types or error conditions, one the device can address automatically and one where the
operator or servie technician is called upon to interact with the unit.
Visualization of the errors detected in the unit is effected via error messages on the serial port, error
screens on the menu display, error texts in the status line or error codes. Error codes are displayed in
the status message and stored in the EEPROM-internal status word.
Error visualization is supported acoustically by a triple signal tone (alarm tone).
Type of failure
Error description
No-error status
0
OK
Error in recipe processing
20
Incomplete recipe found in EEPROM
Error during galvanization.
30
Reference electrode connection error (no galvanization possible)
31
Start position not reached (no liquid container)
32
Cleaning position not reached (no liquid container)
33
Temperature sensor defective
34
Temperature too high during galvanization
35
Temperature too low during heating or galvanization
36
Sample connection error (no galvanization possible)
38
Error in galvanization parameters (general), galvanization halted
39
Error in galvanization parameters (invalid surface)
40
Error in galvanization parameters (invalid process time)
41
Error in galvanization parameters (invalid start current)
42
Error in galvanization parameters (invalid surface) was corrected
Error in data administration (EEPROM error)
50
Invalid recipe counter
51
Error reading data from EEPROM
52
Error writing data to EEPROM
54
EEPROM checksum error
The following is an overview over known unit, operator and process errors. Possible causes and
suggested solutions are shown along with the symptoms of the error.