DeguDent SOLARIS Service Manual
9.2
Commands
Remotely controlling the device using the command interface is restricted to trained
service technicians, since maladjustment of internal parameters can result in
malfunction of the device including adverse effects on the galvanization process.
Service commands can only be issued after the password-protected service mode has
been activated.
9.2.1
Syntax
Commands are issued as plain text. Parameters are separated by commas.
Command names must be entered in lowercase letters.
Decimals are not supported when entering command parameters.
Command parameters must be whole numbers.
CAUTION:
The parameters of all service commands are based on physical channel numbers. These are different
from the logical slot/sample numbers in the local operating menu.
Either type of numbers can be readily converted to the other type:
Channel
(physical number)
Slot
(logical number)
0
2
1
4
2
6
3
8
4
1
5
3
6
5
7
7
This list will be output after issuing the
rc
command.
9.2.2
Standard commands
Command: LS - Service authorization
Command
Parameters
Explanation and examples
i,<password>
Activate service mode (log in)
Example: lsi,<password> -> activates service mode
ls
o
Leave service mode (log out)
Example: lso -> leaves service mode
Command: H – Online help
Command
Parameters
Explanation and examples
hl
-
Online help: Command overview (standard and service commands)
Command: M – Output modes
Command
Parameters
Explanation and examples
0
Mode 0: No output.
1
Mode 1: Output of galvanization parameters during process
2
Mode 2: Comprehensive troubleshooting output
m
4
Mode 4: Minimal output, for internal use only
!