Table of error codes
Occurs on digital
channels
(1 to 32)
Occurs on analog
channels
(41 to 48)
Displayed on the
secondary segments
Cause
Solution
X
„SENSOR E“
Interrupted current loop
or faulty transmitter
Check the transmitter
connection, otherwise
contact the
manufacturer
X
„ NO SENS“
Sensor is not present
(EEPROM is not
communicating)
Disconnect and
reconnect the sensor,
then restart the sensor
by disconnecting and
reconnecting the power
X
„UNK SENS“
Unknows sensor type
Contact the
manufacturer
X
„ELCHEM E“
Type 2 senesor type ID
not found in the table
setting LMP91000
Contact the
manufacturer - FW
update required
X
„SENS CRC“
CRC does not match the
sensor EEPROM
Sensor error - contact
the manufacturer
X
„LMP9 ERR“
LMP91000 does not
respond
Sensor error - contact
the manufacturer
X
„SEN MISM“
The set sensor ID does
not match with the
connected sensor
Checking the
configuration of the
sensor required
X
„ELFAILED“
EL sensor test error
Sensor error, only
informative, subsides by
itself
X
X
„ HEATING“
Transmitter is in
preheating mode
Wait a few minutes, the
transmitter will
automatically enter
measurement mode
X
„EE25 ERR“
Error reading the internal
FLASH
Restart the transmitter.
If the error persists,
contact the
manufacturer
X
„FLASHCRC“
Error reading the internal
FLASH
Restart the transmitter.
If the error persists,
contact the
manufacturer
X
„EEPR CRC“
Error reading the internal
FLASH
Restart the transmitter.
If the error persists,
contact the
manufacturer
X
„INFRA ER“
Infrared sensor error
Replace the sensor
X
„ CALIB18“
The device exceeded the
maximum calibration
interval by 50%
Calibration is necceserry
X
„COMM ERR“
The transmitter doesnt
respond
Check the connection
between the controller
and the transmitter