SETPOINT OPEN ACCESS
If none of the setpoints have their open access option enabled then this feature will
be disabled and the
button will not respond to a short button press. (See 7.2G.)
A
Begin by pressing the
button for less than 3 seconds.
B
The name of the fi rst access-enabled setpoint will appear on the display and
toggle with the current value for that setpoint. Using the
and
buttons,
adjust the selected value. Then press
to accept and continue.
C
The name of the next access-enabled setpoint will appear on the display, along
with its setpoint value. Repeat step 8B. The open access menu will proceed
through all access-enabled setpoints in this fashion. Pressing
for the last ena-
bled setpoint will exit and return to the operational display.
9
RESET PIN NUMBERS / VIEW FIRMWARE VERSION
If you have forgotten your PIN number(s), follow the procedure below to reset both
the Cal and SP PINs to their factory default of 1.
This procedure will also allow you to view the current soft ware installed on your
controller, which may be required for support purposes.
A
Press
,
and
at the same time. (This key combination can be diffi
cult to
execute and you may need several tries to get it right.)
B
A message will appear on the display, with details of the unit's current soft ware
confi guration (Product Name, Firmware Version, and Macro Version). At the
end, you will see
_ _ _ ALL PIN NUMBERS RESET TO 1
.
C
Both the Cal PIN number and the SP PIN number have now been reset to '
1
'.
You can change this, if required, by following the instructions in 6.8 (for the Cal
menu) and 7.3 (for the SP menu), using '
1
' to enter each menu initially.