47
DAC-510E-EN-BG
8 Technical Specifications
8.1 Defibtech DDU-100 AED
8.1.1 Physical
Category
Specification
Size
8.5 x 11.8 x 2.7 inches (22 x 30 x 7 cm)
Weight
Approximately 4.2 lbs (1.9 kg) with DBP-1400 Battery pack
Approximately 4.4 lbs (2 kg) with DBP-2800 Battery pack
8.1.2 Environmental
Category
Specification
Operating / Maintenance
Temperature
0 – 50°C (32 – 122°F)
One Hour Operating
Temperature Limit
(extreme cold)*
-20°C (-4°F)
Humidity
5% – 95% (non-condensing)
Standby / Storage
Temperature
0 – 50°C (32 – 122°F)
Humidity
5% – 95% (non-condensing)
Altitude
-150 to 4500 meters (-500 to 15,000 feet) per
MIL-STD-810F 500.4 Procedure II
Shock / Drop Abuse Tolerance
MIL-STD-810F 516.5 Procedure IV (1 meter, any edge,
corner, or surface, in standby mode)
Vibration
MIL-STD-810F 514.5 Category 20
RTCA/DO-160D, Section 8.8.2, Cat R, Zone 2, Curve G
(Helicopter)
RTCA/DO-160D, Section 8, Cat H, Zone 2, Curves B&R
(Jet Aircraft)
Sealing
IEC 60529 class IP54; Dust Protected, Splash Proof,
(Battery pack installed)
ESD
EN 61000-4-2:1998 Severity Level 4
(Open air discharges up to 8 kV or direct contact
discharges up to 6 kV)
EMC (Emission)
EN 60601-1-2:2001+A1:2006, method EN 55011:1998
Group 1 Level B
(Not to exceed 30 dB μV from 30 Hz to 230 MHz and
not to exceed 37 dB μV from 230 to 1000 MHz
EMC (Immunity)
EN 60601-1-2:2001+A1:2006, method EN 61000-4-
3:1998 Level 3
(Field strength: 10V/m; carrier frequency range:
26 MHz to 1 GHz; AM modulation, 80 percent index,
at 3 frequencies: 1, 5, and 20 Hz)
* From room temperature to temperature extreme, one hour duration.
Содержание Lifeline DCF-100
Страница 2: ......
Страница 3: ...iii DAC 510E EN BG...
Страница 8: ...viii DAC 510E EN BG...
Страница 24: ...16 DAC 510E EN BG...
Страница 36: ...28 DAC 510E EN BG...
Страница 48: ...40 DAC 510E EN BG...
Страница 70: ...62 DAC 510E EN BG...
Страница 72: ...64 DAC 510E EN BG...