7.2 Downloading the Internal Data Log
Regardless of whether a DDC is installed in the unit, select information is recorded internally within
the
DDU-100
AED. The information recorded is limited to:
• The time the AED was turned on.
•
Other data such as event milestones (motion detection, shock advice, shock delivery
information, etc).
•
Eight seconds of ECG data immediately before a shock/no-shock decision, eight seconds
immediately after each shock, and all ECG data during the charging and waiting-to-shock
periods.
•
Note: Audio data is not logged internally.
To download the internally logged information, perform the following procedure:
•
Insert a blank DDC into the unit.
•
Turn the unit on.
•
Once the unit is on, turn it off in data download mode by pushing and holding the ON/OFF
button for at least five seconds.
•
Allow the unit to write the contents of the internal log to the DDC by waiting for the unit to
turn off automatically.
The
DDU-100
will write the contents of the internal log onto the DDC. This information can then
be reviewed using the DefibView software.