31
DAC-510E-EN-BG
If the ASI is flashing red, turn the
DDU-100
AED on. If the unit does not turn on or does not speak,
the AED is non-operational and requires servicing. If the unit does turn on, the voice prompts will
indicate the nature of the problem.
Maintenance Related Voice Prompts:
“Power-on self-test failed, service code ‘xxx’ ”
– This indicates that the
DDU-100
AED
has failed the power-on self-test and is non-operational and needs servicing. The code number will
indicate to the service personnel the type of problem that the unit is experiencing.
“Battery pack self-test failed, service code ‘xxx’ ”
– This indicates that the
DDU-100
AED’s battery pack is non-operational and needs servicing. The code number will indicate to the
service personnel the type of problem that the unit is experiencing.
“Service required”
– This indicates that the unit has detected an error which rendered it
inoperable, and the unit requires servicing.
“Battery pack low”
– This indicates that the battery pack capacity is low and should be replaced
soon. The AED will still be able to deliver at least a minimum of three defibrillation shocks the first
time this message is spoken.
“Replace battery pack”
– This indicates that the battery pack is almost discharged and that the
AED may not be able to deliver defibrillation shocks. The battery pack should be replaced immediately.
“Replace 9 volt battery”
– This indicates that the 9V battery in the battery pack needs to be
replaced. The unit may not provide active status indication during standby mode in this condition,
but the AED is still fully functional when turned on and may be used to treat patients. The 9V
battery should be replaced as soon as possible.
“Pads missing”
– This indicates that pads were not found connected during a self test.
5.2.2 Checking the Condition of the Unit and Accessories
Inspect the unit for cracks or other signs of damage on the case, as well as dirt or contamination,
especially in the areas around the connector socket and battery pack opening.
If any cracks or other signs of damage are observed, remove the unit from service and contact an
autorized service center.
If any dirt or contamination is observed, refer to the “Cleaning” section for guidance on cleaning your unit.