This equipment generates high voltages up
to 35,000V (35kV). Accidental contact with the electrode may result
in an electric shock. Always keep the working end of the electrode
away from your body and never touch it when activated. The
PosiTest HHD
operator should exercise all due care to minimize
their risk of electric shock and be aware of all relevant safety
guidelines before using this equipment. When in
Detection Mode
(pg. 1) generating high voltage or if an arc is being produced, the
PosiTest HHD
will generate broadband Radio Frequency
emissions. Care must be observed in the vicinity of sensitive
electronics or radio equipment.
2
Operational Safety
DeFelsko recommends that the operator read and
fully understand this manual prior to using the
PosiTest HHD
.
IMPORTANT:
WARNING:
For safe
PosiTest HHD
operation:
DO NOT —
w
Use in wet or rainy conditions.
w
Use if you are fitted with a pacemaker.
w
Use without first connecting the ground (earth) cable.
w
Touch or allow anyone else to touch the object being tested.
!
DO —
w
Ensure that all connections are secure and tight.
w
Wear appropriate gloves and footwear
w
Use flagging tape and signage to keep other personnel away
from the testing area.
w
Communicate the hazards associated with holiday testing with
all personnel on-site via tailgate meetings or Field Level
Hazard Assessments.
w
Power OFF the detector before touching the spring or brush.
w
Confirm the testing voltage requirements and ensure this is
communicated to all personnel involved or working in the area.
Содержание PosiTest HHD
Страница 1: ...Instruction Manual v 1 0 High voltage Holiday Detector ...
Страница 2: ......
Страница 19: ...Manual de instrucciones v 1 0 High voltage Holiday Detector Detector de Porosidad de Alto Voltaje ...
Страница 36: ...Gebrauchsanweisung v 1 0 High voltage Holiday Detector Hochspannungs Porenprüfgerät ...
Страница 53: ...Notice d instruction v 1 0 High voltage Holiday Detector Détecteur de Micro fissure Haute tension ...
Страница 71: ......