14
PosiTest HHD turns OFF while in use
w
Check Battery:
A low or defective battery will not be able to
sustain the power required to operate the
PosiTest HHD
. Charge
or replace the battery. To determine if the battery is defective
refer to
Battery Charging
(pg. 12).
When generating high voltage, the PosiTest HHD display is
lower than the test voltage.
w
The
PosiTest HHD
integrated jeep/crest meter ALWAYS displays
the actual voltage being generated. Although the
PosiTest HHD
will regulate the output voltage to match the test voltage, there
are circumstances when it is not possible to generate the test
voltage. For example, overloading of the electrode by setting the
voltage higher than recommended for the coating system under
inspection may result in a generated voltage lower than the test
voltage. Reducing the brush or electrode size will also reduce
loading.
Accessories
A variety of optional accessories are available to help you get the
most out of your
PosiTest HHD
.
w
Handheld Wand w/ 1.5m (5’) Insulated Cable
w
10cm (4”) Conductive Rubber Paddle
w
Steel Flat Wire Brushes
w
Steel Rolling Spring Electrodes
w
Rolling Spring Electrode Connector
w
Rolling Spring Coupler (sold in pairs)
w
15cm (6”) Extension Rod for use in areas of limited
clearance/headroom
w
PosiTest HHD Verifier
(Calibrated peak reading voltmeter can
be used to verify operation of the
PosiTest HHD
.)
w
Tinker & Rasor Electrode Adaptor
w
PCWI Electrode Adaptor
For a complete list of accessories see: www.defelsko.com/hhd
Содержание PosiTest HHD
Страница 1: ...Instruction Manual v 1 0 High voltage Holiday Detector ...
Страница 2: ......
Страница 19: ...Manual de instrucciones v 1 0 High voltage Holiday Detector Detector de Porosidad de Alto Voltaje ...
Страница 36: ...Gebrauchsanweisung v 1 0 High voltage Holiday Detector Hochspannungs Porenprüfgerät ...
Страница 53: ...Notice d instruction v 1 0 High voltage Holiday Detector Détecteur de Micro fissure Haute tension ...
Страница 71: ......