Stops the recording process, closes the current batch, and
removes batch information from the display.
Setup Menu
Pull Parameters
Dolly Size:
Select the dolly size being utilized for the test (10,
14, 20, 50, or 50 x 50 mm)
Rate:
User selectable pull rate (unit/sec). Specified rate is
maintained for the duration of the test.
Limit:
Pressure will build-up until the set
Limit
is reached or the
dolly is pulled from the surface, whichever occurs first. By default,
the instrument will pull to the maximum limit for selected dolly size
(see
Technical Data
pg.16).
Select desired parameter using the
buttons or touch
screen. Use the
and buttons to adjust.
Delete
Removes a batch and associated measurements from memory.
Hold Time:
(up to 60 seconds)
pressure is
maintained
(held)
at defined
Limit
.
Default: 0 seconds
Hold Time
is
displayed on
the pull chart
(shaded area)
Defined
Hold Time
Orientation
Rotates display between landscape
(default)
or portrait views.
Ideal for use in horizontal or vertical positions.