5. INSTALLING THE DETECTOR SYSTEM
5.1. Transport Considerations
Warning
#2
Avoid vibration and shock when moving the detector.
Caution
#9
Use the included transport hook for transporting the detector where applicable.
The detector has a transport hook which has to be used for lifting the detector. When using the transport hook, make sure it
is properly tightened. Only use an appropriate lifting device with a carrying capacity of at least 150 kg to move the detector.
Make sure two persons always move the detector. When moving the detector, make sure the protection cover is mounted
and all cables and hoses are disconnected.
The detector has been delivered in a robust transport box. Please keep this transport box for transport or storage purpose.
5.2. Mounting
Do not place the detector near heat sources or in a place subject to direct sunlight, excessive dust or mechanical shock.
Make sure that the detector has adequate ventilation.
• Do not cover any air intakes or outlets.
• Place the detector in a location with adequate air circulation.
• Make sure the detector has enough space for proper ventilation (minimum wall distance: 170 mm).
• Do not operate the detector in a closed environment.
The detector can be mounted in the ways which are described below.
5.2.1. Mounting from Above
Use the mounting brackets as depicted in figure 5.1. These mounting brackets have to be mounted on the base plate of
the detector. The detector should be mounted using all four outer 7 mm holes (indicated with yellow circles).
Caution
#10
Make sure the mounting brackets are mounted and properly tightened using the screws provided.