Table 4.3:
Air Connectors
Connector
Description
N
2
/Dry Air
Nitrogen or dry air for humidity control. Use a hose/pipe with an
outer diameter of 6 mm.
To release the tube press and hold the blue ring before pulling at
the tube.
Table 4.4:
Coolant Connectors
Connector
Description
IN
Coolant inlet
OUT
Coolant outlet
Caution
#6
Use only the included thermal stabilization unit.
4.2. Detector Control Unit
4.2.1. Configuration of the Detector Control Unit
Please do not install or run any other software on the computer, except tools and software which are necessary for config-
uring your data acquisition protocol.
The detector control unit is set up with a standard installation of the CentOS 6.x distribution. Regular system updates can
be made. However, to avoid operational deterioration do not update the system while the detector is taking data.
Caution
#7
Do not remove the symbolic link in the directory
~/p2_det/images
, which points to the
images
directory.
The detector control unit has to be connected point-to-point to the detector via 2 x 10GBase-T Ethernet. The detector
control unit can be integrated into the site network infrastructure using one of the interfaces described in section 4.2.2. The
detector control unit is optimised for performance and stability of operation. In order to achieve these goals we deliver the
detector control unit with fixed firmware (bios etc.) and software (OS) version.
Figure 4.5:
PILATUS3 S 6M detector control unit as seen from the front.
Caution
#8
Pushing the power button on the front panel longer than 2 seconds will immediately halt the detector control
unit. All image data on the detector control unit will be permanently lost.