Technical_Specification_PILATUS_1M_V1_7
26/31
7
Getting Started
Before operating the detector, make sure you have read the previous chapters
in the technical specification and the user manual.
Check these items before turning the detector system on:
Mount the detector properly
Connect the detector to power
Connect the detector to nitrogen or dry air at the recommended flow
rate
Connect the water tubes. Make sure they are properly mounted on both
sides
Connect the PC and the detector data cables
Attach monitor and keyboard to the computer
7.1 Startup Sequence
Turn on Nitrogen or dry air flow at least 30 minutes before turning on
the detector
Turn on the PC
Start a shell
The default path is: /home/det
Change the directory to: p2_det
Turn on the power switche at the back-side of the detector
Turn on the cooling unit at the same time as the detector is started
Type runtvx
Runtvx starts a script file which initializes the detector system and
opens the Camserver and TVX windows