![Dectris PILATUS 1M Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/dectris/pilatus-1m/pilatus-1m_operating-procedures-manual_2462220020.webp)
Technical_Specification_PILATUS_1M_V1_7
20/31
5.2.2 Mounting from Below
From below with four M6 bolts.
Figure 10. Mounting from below, bottom view
The four M6 bolts should not intrude into the detector more than 16
mm.
Make sure the detector is properly mounted.
Make sure the detector has enough space for proper ventilation
(minimum wall distance: 170 mm). Do not operate the detector in a closed
environment.
Mounting points when mounting from the bottom