Technical_Specification_PILATUS_1M_V1_7
14/31
Connector
Description
Ground
Functional ground of the detector system.
Although the detector might be grounded via
the mounting bolts, the detector can be grounded
additionally via the functional ground connector at the
back (M6 screw-in tap hole) to establish a defined
grounding.
N2
Nitrogen for humidity control.
For details see section 5.4.
Pipe:
Use a pipe with outer diameter of 6 mm and an inner
diameter of 4 mm (e.g: Legris Type 1025U06).
Di-H2O
Use deionized water for cooling.
Pipes:
Use for inlet and outlet a pipe with an inner diameter of 9
mm.
The fixation pieces and the transport hook are stored on the back of the
detector.