Set damping time
8
Density (to be used)
9
Arbitrary point migration input compensation value, compensation
value = P actual - P pre-output
11
LRV calibration
12
URV calibration
17
View Sensor Values
Calibration of the instrument can be done as described in "7.1.4 Data Setting Method", with upper
and lower range limits and damping settings.
7.1.6 Unit Setting
The units that can be set are: m, cm, mm, feet, and inch.
① Press and hold the '←┘' to enter the configuration mode (the No. 1 menu will be displayed first,
and the up and down arrows on the left will be lit).
② Press the "↑" to switch the menu to No.2 and display the current unit symbol in the lower right
corner.
Unit Setting
③ Press the '←┘' key again, the unit starts flashing units, now press the '↑' key or '↓' to switch to
the unit you want to set.
④ Press "←┘" again, the unit is set and the up and down arrows on the left side are lit.
7.1.7 Range Setting
Set the lower range LRV
① Press and hold the '←┘' key to enter the configuration mode (the No. 1 menu will be displayed
first, and the up and down arrows on the left will be lit).
② Press the "↑" key to switch the menu to No.3, and the LCD will show the current LRV value.