CAL3K-A INSTALLATION GUIDE V1.1  |  

www.ddscalorimeters.com

  |  [email protected] 

 

 

 

Please see above for the dimensions, weights and power requirements per unit.  The BALANCE data 

would be dependent on the make and model that you will be using. 

 
 

SAFETY PRECAUTIONS 

CALORIMETER (3K-1_A) 

 

Electrical Shock: There are no user serviceable parts inside. 

 

Do not open the bottom panel unless the supply cable has been disconnected. 

 

Dangerous voltages will be present inside the unit after the supply cable has been disconnected. 

 

 

DO NOT TAMPER WITH THE EARTHING - CONSULT A QUALIFIED ELECTRICIAN SO THAT THE 

EARTHING CAN BE CHECKED. 

 

DO NOT USE ANY OTHER PIPE OTHER THAN THE HIGH PRESSURE PIPE SUPPLIED FOR OXYGEN 

FILLING. 

 

 

NOTE: With some applications you may need to fill the Vessel to less than 2400KPA e.g.1500KPA (15 bar), 

however this is only applicable when you have a sample which is highly combustible, for example with wood 

related samples. THE FILLING PRESSURE IS ADJUSTABLE on the regulator, the filling speed is adjustable on 

the filling station. 

 

 

 

REMEMBER: You may fill the Vessel LESS than 3000KPA but NEVER more than 3000KPA. 

 

FILLING STATION (CAL2K-3) 

 

Secure the Filling Station to the table. 

 

Install the flexible high pressure pipe supplied without any sharp bends. 

 

Check the nylon high pressure pipe supplied for any mechanical damage before putting it under pressure. 

 

Secure the pipe nuts to the pressure regulator side, then clamp the ferrule to the high pressure pipe on the 

gauge block side. 

 

If the oxygen bottle is not situated close to the Filling Station or if it is in another room or perhaps outside, 

then a solid, permanent installation of pressure piping should be installed, as well as a shut off valve and 

pressure gauge within easy reach of the CAL2K-3 Filling Station. 

 

If bottled oxygen is used, a suitable High Pressure oxygen regulator must be supplied by your Agent, or 

sourced locally. Ordering this item from the factory ensures that the necessary modifications have been 

made to fit the CAL2K-3 Filling Station. 

 

The oxygen bottle should be secured according to regulations. 

 

DO NOT USE ANY OTHER PIPE OTHER THAN THE HIGH PRESSURE PIPE SUPPLIED. 

 

 

Содержание CAL3K-A

Страница 1: ...cturers of CAL2K CAL3K Oxygen Bomb Calorimeters CAL3K A INSTALLATION GUIDE V1 1 www ddscalorimeters com calo ddsystems co za CAL3K A INSTALLATION GUIDE VERSION 1 1 MANUFACTURED BY DIGITAL DATA SYSTEMS...

Страница 2: ...A CALORIMETER WITH EXTERNAL oxygen filling 6 THE 3K 2 AIR COOLER 6 THE 3K 5 BAYONET VESSEL 7 THE CAL2K 3 OXYGEN FILLING STATION 7 7 UNPACKING AND INSPECTION 7 SETTING UP 8 PREREQUISITES 8 SITE SELECT...

Страница 3: ...INATE USER TEST 16 16 CAL3K A USER TEST 16 16 SUBJECT HELP 16 OTHER KEYBOARD COMMANDS 16 ABORT 16 FIRE MANUAL 16 CALIBRATION 16 READ RESULT 17 CONNECTING TO A BALANCE 17 17 CALIBRATION 18 INTELCAL CAL...

Страница 4: ...and the factory calibration But there is more Information and capability This is not explained here but in an ON LINE version It is safe to say that you need a PC or laptop to change from the basic o...

Страница 5: ...Bottle 2 3K A INST CAL3K A Installation Guide Printed 1 3K 2 CAL3K Air Cooler 1 3K 2 KT CAL3K Air Cooler Installation Kit 1 Included in Kit 3K SM Power Supply 1 KETTLECORD Mains Cable 220 V 1 CAL2K 3...

Страница 6: ...e calorimeter and the lid is closed The rest is automatic Width 270mm Height 200mm Depth 380mm Weight 8kg Power 220 240V 12W THE 3K 2 AIR COOLER The air cooler requires no water piping or other resour...

Страница 7: ...ight 150mm Weight 1 9kg THE CAL2K 3 OXYGEN FILLING STATION The oxygen filling station is connected to the oxygen bottle via a regulator The regulator reduces the bottle pressure to 30bar which is adju...

Страница 8: ...contact your agent SETTING UP PREREQUISITES Prior to installation of the CAL3K A System the User must provide the following 1 A mains outlet for the 3K 1_A Calorimeter and a mains outlet for the 3K 2...

Страница 9: ...HE FILLING PRESSURE IS ADJUSTABLE on the regulator the filling speed is adjustable on the filling station REMEMBER You may fill the Vessel LESS than 3000KPA but NEVER more than 3000KPA FILLING STATION...

Страница 10: ...to the outside or any suitable scrubber to prevent the burned gas to enter the room COOLER 3K 2 Electrical shock There are no user serviceable parts inside VESSEL 3K 5 The Vessel has been pressure tes...

Страница 11: ...DS CALORIMETERS www ddscalorimeters com Manufacturers of CAL2K CAL3K Oxygen Bomb Calorimeters CAL3K A INSTALLATION GUIDE V1 1 www ddscalorimeters com calo ddsystems co za CAL3K A REAR PANEL CONNECTION...

Страница 12: ...test is logged into the Event Logging and the User Test Log It is recommended to repeat the hardware once a day or once per shift The hardware test takes approximately 1 minute to complete NOTE While...

Страница 13: ...CAL3K A hardware test The unit cycles through the following steps and stops for operator action or when a fault is encountered Will stop the test The results of the test are logged together with the V...

Страница 14: ...password Function keys They serve as a shortcut to the keyboard commands The following function keys are used F1 Mass entry F1 Enter 0 xxxx Enter F2 Sample Identification F2 Enter mysampleId Enter F3...

Страница 15: ...6 character or shorter string SPIKE MASS Enter the weight of the spike mass here A spike mass of zero disables spiking For details of spiking see www ddscalorimeters com BALANCE MASS The supplied bala...

Страница 16: ...nit waits for an action a more detailed subject help is available to explain what is needed Press whenever something does not progress and read the comments OTHER KEYBOARD COMMANDS Here are some handy...

Страница 17: ...e stable balance weight should show on the display If you press Enter then the mass is locked See MASS ENTRY to unlock it PERFORM THE FIRST SAMPLE The object of this exercise is to check that everythi...

Страница 18: ...e are many additional steps and procedures that you can do whilst running an actual CV sample You will find these How to Steps via the On line Help Guide available on the website www ddscalorimeters c...

Страница 19: ...rocedure with the second vessel The calibration data set is stored in the vessel calibration history for the chassis number ROUTINE SAMPLE DETERMINATION The procedure is the same If you perform multip...

Страница 20: ...be binding on dds unless recorded in writing and signed by a duly authorised officer of the company dds agrees to repair and replace with a new or remanufactured part any part other than those exclud...

Страница 21: ...hat is sent to dds for Repair is tested and checked and leaves dds s Factory in 100 working order It is the Owner s responsibility to inspect any repaired product on receipt thereof and confirm in wri...

Страница 22: ...AME OWNER S CUSTOMER S COMPANY NAME AUTHORISED BY INSTALLED AT TEL NO FAX NO CONTACT PERSON INSTALLED BY DATE INSTALLED REPRESENTATIVE OF DEALER AGENT CUSTOMER ORDER NUMBER AND DATE INSTALLATION DETAI...

Отзывы: