CAL3K-A INSTALLATION GUIDE V1.1  |  

www.ddscalorimeters.com

  |  [email protected] 

READ RESULT 

This command is required to read the previous stored results. The ‘last one first’ result can be retrieved by: 

 

  

READ   

   

 

 

The unit displays the Last result number and the result details in line #4. 

 

Press (Enter)   

 

The unit displays the previous result 

 

Press (Enter)   

 

repeat until you have all results! 

 

 

 

Note: The results can only be cleared from a PC or mobile device if allowed to do so! 

 

CONNECTING TO A BALANCE 

This step is not absolutely necessary. The weight (mass) of the sample can be transferred manually from the balance 

to the CAL3K-A and entered on the keyboard. The balance connection cable is included in the kit. If the unit is 

installed near the balance then connect up. The default balance speed is 2400baud. Set your balance to this speed! 

The balance speed can be changed from the PC only. Test the connection by pressing (F5). Then the stable balance 

weight should show on the display. If you press (Enter) then the mass is locked. See MASS ENTRY to unlock it. 

 
 

 

PERFORM THE FIRST SAMPLE 

The object of this exercise is to check that everything functions.  We assume that the CAL3K-A and the Vessel 

hardware has been tested and found to be OK. 

 

The sample test requires the following: 

 

The default setup (set by the factory) 

 

One Benzoic Acid (CAL2K-4-BA) tablet, supplied in 3K-1 Installation kit 

 

Firing cotton (3K-5-FC), supplied in 3K-5 Installation Kit 

Firing wire (3K-5-FW), (already fitted to vessel lid assembly, also supplied in 3K-5 Installation kit), check that 

this is present 

 

Minimum fuss, Minimum explanations, Minimum keyboard enter 

 

Herewith the operating sequence: 

 

1.

 

Weigh a Benzoic Acid (supplied) on a scale to 4 decimal places with the tarred crucible. Do not touch the 

tablet but use the supplied tweezers. 

2.

 

Prepare a vessel: Attach Firing cotton thread to the Firing wire. Insert the crucible with the BA tablet in to 

the holder, and position the tabled on top of the cotton thread. Handle the lid gently because if the tabled 

moves and does not touch the cotton thread it is unlikely that the tabled fires. Then gently close the vessel. 

Do not over tighten.  

Содержание CAL3K-A

Страница 1: ...cturers of CAL2K CAL3K Oxygen Bomb Calorimeters CAL3K A INSTALLATION GUIDE V1 1 www ddscalorimeters com calo ddsystems co za CAL3K A INSTALLATION GUIDE VERSION 1 1 MANUFACTURED BY DIGITAL DATA SYSTEMS...

Страница 2: ...A CALORIMETER WITH EXTERNAL oxygen filling 6 THE 3K 2 AIR COOLER 6 THE 3K 5 BAYONET VESSEL 7 THE CAL2K 3 OXYGEN FILLING STATION 7 7 UNPACKING AND INSPECTION 7 SETTING UP 8 PREREQUISITES 8 SITE SELECT...

Страница 3: ...INATE USER TEST 16 16 CAL3K A USER TEST 16 16 SUBJECT HELP 16 OTHER KEYBOARD COMMANDS 16 ABORT 16 FIRE MANUAL 16 CALIBRATION 16 READ RESULT 17 CONNECTING TO A BALANCE 17 17 CALIBRATION 18 INTELCAL CAL...

Страница 4: ...and the factory calibration But there is more Information and capability This is not explained here but in an ON LINE version It is safe to say that you need a PC or laptop to change from the basic o...

Страница 5: ...Bottle 2 3K A INST CAL3K A Installation Guide Printed 1 3K 2 CAL3K Air Cooler 1 3K 2 KT CAL3K Air Cooler Installation Kit 1 Included in Kit 3K SM Power Supply 1 KETTLECORD Mains Cable 220 V 1 CAL2K 3...

Страница 6: ...e calorimeter and the lid is closed The rest is automatic Width 270mm Height 200mm Depth 380mm Weight 8kg Power 220 240V 12W THE 3K 2 AIR COOLER The air cooler requires no water piping or other resour...

Страница 7: ...ight 150mm Weight 1 9kg THE CAL2K 3 OXYGEN FILLING STATION The oxygen filling station is connected to the oxygen bottle via a regulator The regulator reduces the bottle pressure to 30bar which is adju...

Страница 8: ...contact your agent SETTING UP PREREQUISITES Prior to installation of the CAL3K A System the User must provide the following 1 A mains outlet for the 3K 1_A Calorimeter and a mains outlet for the 3K 2...

Страница 9: ...HE FILLING PRESSURE IS ADJUSTABLE on the regulator the filling speed is adjustable on the filling station REMEMBER You may fill the Vessel LESS than 3000KPA but NEVER more than 3000KPA FILLING STATION...

Страница 10: ...to the outside or any suitable scrubber to prevent the burned gas to enter the room COOLER 3K 2 Electrical shock There are no user serviceable parts inside VESSEL 3K 5 The Vessel has been pressure tes...

Страница 11: ...DS CALORIMETERS www ddscalorimeters com Manufacturers of CAL2K CAL3K Oxygen Bomb Calorimeters CAL3K A INSTALLATION GUIDE V1 1 www ddscalorimeters com calo ddsystems co za CAL3K A REAR PANEL CONNECTION...

Страница 12: ...test is logged into the Event Logging and the User Test Log It is recommended to repeat the hardware once a day or once per shift The hardware test takes approximately 1 minute to complete NOTE While...

Страница 13: ...CAL3K A hardware test The unit cycles through the following steps and stops for operator action or when a fault is encountered Will stop the test The results of the test are logged together with the V...

Страница 14: ...password Function keys They serve as a shortcut to the keyboard commands The following function keys are used F1 Mass entry F1 Enter 0 xxxx Enter F2 Sample Identification F2 Enter mysampleId Enter F3...

Страница 15: ...6 character or shorter string SPIKE MASS Enter the weight of the spike mass here A spike mass of zero disables spiking For details of spiking see www ddscalorimeters com BALANCE MASS The supplied bala...

Страница 16: ...nit waits for an action a more detailed subject help is available to explain what is needed Press whenever something does not progress and read the comments OTHER KEYBOARD COMMANDS Here are some handy...

Страница 17: ...e stable balance weight should show on the display If you press Enter then the mass is locked See MASS ENTRY to unlock it PERFORM THE FIRST SAMPLE The object of this exercise is to check that everythi...

Страница 18: ...e are many additional steps and procedures that you can do whilst running an actual CV sample You will find these How to Steps via the On line Help Guide available on the website www ddscalorimeters c...

Страница 19: ...rocedure with the second vessel The calibration data set is stored in the vessel calibration history for the chassis number ROUTINE SAMPLE DETERMINATION The procedure is the same If you perform multip...

Страница 20: ...be binding on dds unless recorded in writing and signed by a duly authorised officer of the company dds agrees to repair and replace with a new or remanufactured part any part other than those exclud...

Страница 21: ...hat is sent to dds for Repair is tested and checked and leaves dds s Factory in 100 working order It is the Owner s responsibility to inspect any repaired product on receipt thereof and confirm in wri...

Страница 22: ...AME OWNER S CUSTOMER S COMPANY NAME AUTHORISED BY INSTALLED AT TEL NO FAX NO CONTACT PERSON INSTALLED BY DATE INSTALLED REPRESENTATIVE OF DEALER AGENT CUSTOMER ORDER NUMBER AND DATE INSTALLATION DETAI...

Отзывы: