4.9
Principle of Agent Identification
The anaesthetic agent identification bench identifies halothane,
enflurane, isoflurane, desflurane and sevoflurane.
The operation of the bench is based on infrared absorption at 3.3 um
range. It measures the spectrum of the gas between 3.24 um and 3.39
um. Because the spectrum of each of the anaesthetic agents is different
it is possible to identify them.
The bench consists of an infrared source, a measuring chamber, a
rotating filter and a detector. The peak wavelength of the narrow
bandpass filter changes when the angle between the light path and the
filter is changed. When the filter rotates the required spectrum is
scanned through.
The agent or a mixture of agents is calculated from the measured
spectrum using stored reference spectrums of each agent.
Figure 4.8 shows the absorption spectra of anaesthetic agents.
Figure 4.8
Anaesthetic Agents Gas Absorption Spectra
January 15th, 1997/6
Содержание CAPNOMAC ULTIMA
Страница 38: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 4 13...
Страница 39: ...4 14 January 15th 1997 6...
Страница 46: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 4 21...
Страница 51: ......
Страница 53: ......
Страница 57: ......
Страница 58: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 5 10...
Страница 60: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 82: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 5 34 January 15th 1997 6...
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 108: ......
Страница 109: ......
Страница 111: ......
Страница 113: ...CAPNOMAC ULTIMA TM SERVICE MANUAL Preamplifier Figure 5 29 ASX Agent identification benches January 15th 1997 6...
Страница 117: ...ASX Preamplifier Board...
Страница 119: ......
Страница 120: ......
Страница 121: ......
Страница 122: ......
Страница 143: ......
Страница 151: ......
Страница 152: ......
Страница 153: ......
Страница 159: ......
Страница 165: ......
Страница 166: ......
Страница 183: ...CAPNOMAC ULTIMA TM SERVICE MANUAL Other external parts January 15th 1997 6...
Страница 185: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 9 6 January 15th 1997 6...
Страница 187: ...CAPNOMAC ULTIMATM SERVICE MANUAL 9 8 Figure 9 1 Exploded picture of the monitor all except 27 model 0 7...
Страница 188: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 9 9 Figure 9 2 Exploded picture of the monitor 27 model revisions 08 and lower...
Страница 189: ...r 71 I I r a...
Страница 190: ......
Страница 206: ...DATEX CAPNOMAC ULTIMATM SERVICE MANUAL Figure 5 17 Video control board block diagram May 2nd 1991 1...
Страница 211: ...DATEX CAPNOMAC ULTIMA SERVICE MANUAL 5 34 November 1st 1993 4...
Страница 213: ...DATEX CAPNOMAC ULTIMA SERVICE MANUAL 5 36 Figure 5 20 Power supply board block diagram May 2nd 1991 1...
Страница 214: ......
Страница 217: ......
Страница 218: ......
Страница 226: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 12 8 Status code definition cont 1st digit 7 January 15th 1997 6...
Страница 231: ...CAPNOMAC ULTIMA SERVICE MANUAL 12 13 January 15th 1997 6...
Страница 232: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 1 2 1 4 Leak Disconn Obstr Fuite Deconn Obstr Leck Diskonn Obstr January 15th 1997 6...