8 - l
8
FUNCTIONAL FIELD CHECK PROCEDURE
8.1
Preoperative check list
Perform the following tests daily to assure proper operation of the
monitor.
WARNING:
If the monitor fails to respond as described, stop
using the monitor, troubleshoot, and correct the situation.
1.
AIRWAY GASES
Confirm that the water trap container is empty and tightly
attached, and the water trap is properly attached to the
monitor.
Connect the gas sampling line to the connector on the water
trap.
NOTE:
Occlusion pressure limit is set during start-up. Attach
sampling line before turning monitor on.
Turn the power on. The monitor performs a start-up and
self test procedure. Wait until message “CALIBRATING
GAS SENSOR” disappears.
Confirm that the inspiratory O
2
reading is about 21 %, and
no other gas is detected.
Breath five times into the sampling line to verify that a
capnogram and end tidal value are displayed and they are
in expected range. If not, calibrate the monitor, see the
Operator’s Manual. Calibrate the monitor every 6 months
even if there is no sign of errors.
Stop breathing into the sampling line. After 20 seconds the
monitor should give the first audible and visible alarm
indicators.
January 15th, 1997/6
Содержание CAPNOMAC ULTIMA
Страница 38: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 4 13...
Страница 39: ...4 14 January 15th 1997 6...
Страница 46: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 4 21...
Страница 51: ......
Страница 53: ......
Страница 57: ......
Страница 58: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 5 10...
Страница 60: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 82: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 5 34 January 15th 1997 6...
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 90: ......
Страница 91: ......
Страница 92: ......
Страница 93: ......
Страница 108: ......
Страница 109: ......
Страница 111: ......
Страница 113: ...CAPNOMAC ULTIMA TM SERVICE MANUAL Preamplifier Figure 5 29 ASX Agent identification benches January 15th 1997 6...
Страница 117: ...ASX Preamplifier Board...
Страница 119: ......
Страница 120: ......
Страница 121: ......
Страница 122: ......
Страница 143: ......
Страница 151: ......
Страница 152: ......
Страница 153: ......
Страница 159: ......
Страница 165: ......
Страница 166: ......
Страница 183: ...CAPNOMAC ULTIMA TM SERVICE MANUAL Other external parts January 15th 1997 6...
Страница 185: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 9 6 January 15th 1997 6...
Страница 187: ...CAPNOMAC ULTIMATM SERVICE MANUAL 9 8 Figure 9 1 Exploded picture of the monitor all except 27 model 0 7...
Страница 188: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 9 9 Figure 9 2 Exploded picture of the monitor 27 model revisions 08 and lower...
Страница 189: ...r 71 I I r a...
Страница 190: ......
Страница 206: ...DATEX CAPNOMAC ULTIMATM SERVICE MANUAL Figure 5 17 Video control board block diagram May 2nd 1991 1...
Страница 211: ...DATEX CAPNOMAC ULTIMA SERVICE MANUAL 5 34 November 1st 1993 4...
Страница 213: ...DATEX CAPNOMAC ULTIMA SERVICE MANUAL 5 36 Figure 5 20 Power supply board block diagram May 2nd 1991 1...
Страница 214: ......
Страница 217: ......
Страница 218: ......
Страница 226: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 12 8 Status code definition cont 1st digit 7 January 15th 1997 6...
Страница 231: ...CAPNOMAC ULTIMA SERVICE MANUAL 12 13 January 15th 1997 6...
Страница 232: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 1 2 1 4 Leak Disconn Obstr Fuite Deconn Obstr Leck Diskonn Obstr January 15th 1997 6...