S/5 Anesthesia Monitor and S/5 Critical Care Monitor
Document No. 8002948
28
5.8
Record Data
In this menu the user can specify where to print from any
service menu. The setting goes into the permanent
memory of the monitor. Only one option can be chosen at
a time. The options where to print are:
To Printer, To
Recorder
and
To Memory Card
. Only the option
To
Recorder
can be selected currently. When any of the
front panel keys of the recorder module, M-REC is
pressed the output is the recorder module despite of the
setting of Record Data menu.
5.9
Remote Access
Remote Access
menu is for setting the interface with an
external PC that runs the Virtual Support
TM
software,
which is a service diagnostic tool for remote diagnostics
purposes.
When the access mode is set to
Normal
(the default), the
PC can only read service menu data from the monitor.
With the access mode
Extended
the PC can also activate
the monitor’s service menu functions.
NOTE: Patient monitoring is not possible when
the access mode Extended is selected. The
access mode turns back Normal if the monitor
is restarted or if the service menu is closed.