CPU boards and softwares
Document No.8002948
11
!
3.
Insert the software card into the card drive slot and firmly press the card in position.
4.
Turn the power on.
5.
Wait for approximately 80 seconds. After the start up screen appears enter the Service View
and make sure that the information regarding monitor software has been updated. Memorize
the serial number of new software.
6.
Remove the software card and attach the lid.
7.
Pick up the software device plate with the serial number of the new software and attach the
device plate on the B-CPU5/B-CPU4 rear panel.
NOTE: The license agreement, if delivered with the monitor software, needs to be in accordance
with the corresponding monitor software serial number. Make sure you archive the license
agreement in a secure location.
8.
Perform Factory Reset.
9.
Set the time and date.
10.
Set the monitor’s network communication according to the used network software, if
necessary.
Network –service menu:
Network software S-CNET99 -> DRI Level = 1999
Network software S-CNET01 -> DRI Level = 2001
NOTE: If the DRI level is changed, the monitor will restart automatically.
11.
Check that there are no error messages on the screen.
12.
Restore the original user settings, if necessary.
See the troubleshooting flowchart for software card, if the downloading of new software failed.
NOTE: Right after the monitor software downloading the start-up time is considerably longer.