92
Standard Calibration
During this process, the printhead assembly can be raised for visual access. In addition, active
sensor readings can be used to further determine the best position for small, position-critical
TOF notches or marks. Three readings are required:
•
Empty: A measurement taken with no media over the sensor.
•
TOF Indicator: A measurement taken with only the media backing (liner) material,
notch, or reflective mark placed over the sensor.
•
Paper: A measurement taken with the label (and the liner) placed over the sensor.
Before calibrating, ensure that the Advanced Menu system is enabled, and that the Media
Sensor Type has been selected. Then, use the steps below and the displayed prompts to
guide you through the procedure:
Step
Action
Displayed
Message
Comment(s)
1
Press the MENU Button.
CANNOT CALIBRATE
-or-
UNCALIBRATED
See MENU ORGANIZATION for
menu navigation tips.
2
Press the ENTER Key to
access MEDIA SETTINGS, and
then use the UP Button to scroll
to SENSOR CALIBRATION.
MEDIA SETTINGS
This is the MEDIA SETTINGS menu
branch.
3
Press the ENTER Key to
access SENSOR
CALIBRATION. Press the
ENTER Key again, and then
press the YES Key to proceed.
PERFORM
CALIBRATION
You are beginning the calibration
procedure. Press the NO Key to abort
this procedure.
4
Ensure that no label stock is
installed in the printer then
press the ESC Key.
REMOVE LABEL
STOCK
PRESS ESC KEY
yyy
This sets the empty value - where
‘yyy’ is a numerical value representing
the sensor reading when no media is
installed.
Содержание H Class
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 6: ......
Страница 12: ...4 ...
Страница 23: ...15 Network Status page TCP IP Configuration page ...
Страница 42: ...34 ...
Страница 142: ...134 ...
Страница 144: ...136 ...
Страница 152: ...144 ...
Страница 156: ...148 ...
Страница 160: ...152 ...