82
H-Class
Advanced Entry Calibration (continued)
Step
Action
Displayed Message
Comment
F
Use the buttons to set the
Gain Number to 00 and
then press ENTER.
Record the sensor reading
as a Label Value for Gain
Number 00 in a table (32
rows by four columns,
with headings similar to
those shown below.)
TRAN SENSOR GAIN
yyy (0 - 31)
00
This is the Label Value for
a gain setting of 00,
where ‘yyy’ represents
the current sensor
reading.
Sample Calibration Table
Gain Number
Label Value
TOF Value
Difference Value
00 252
01
02
…
31
Step
Action
Displayed Message
Comment
G
Use the buttons to
increment the Gain
Number by one and then
press ENTER. Record the
Label Value. Repeat this
process for each Gain
Number.
TRAN SENSOR GAIN
yyy (0 - 31)
01
These are Label Values,
where ‘yyy’ represents
the current sensor
reading.
Sample Calibration Table
Gain Number
Label Value
TOF Value
Difference Value
00 252
01 250
02 248
… …
31 09
Содержание H-4212
Страница 1: ...Operator s Manual ...
Страница 2: ......
Страница 5: ......
Страница 40: ...30 H Class ...
Страница 84: ...74 H Class ...
Страница 144: ...134 H Class ...
Страница 148: ...138 H Class ...
Страница 156: ...146 H Class ...
Страница 160: ...150 H Class Network Status page TCP IP Configuration page ...
Страница 162: ...152 H Class ...
Страница 163: ...H Class 153 Printer Options Communications and Diagnostics pages ...
Страница 164: ...154 H Class Network Print Options page ...
Страница 165: ...H Class 155 Reset Network Parameters page Change Password page ...
Страница 166: ...156 H Class ...
Страница 180: ...170 H Class ...
Страница 186: ...176 H Class ...