TROUBLESHOOTING
99
9
TROUBLESHOOTING GUIDE
Problem
Suggestions
Reading
:
Not possible to read the
target barcode (always
returns No Read) or the
Auto Setup procedure
Fails.
Check synchronization of reading pulse with object to read:
Is the scan line correctly positioned?
Place barcode in the center of scan line and run Test mode
(selectable by Genius™ as an Operating Mode). If you still have
trouble, check the following:
Is the reading distance within that allowed (see reading
diagrams)?
Is the Tilt angle too large?
Is the Skew angle less than 10° (direct reflection)?
Choose the Code Definition branch and enable different
Code Symbologies (except Pharmacode). Length = Min
and Max (variable).
Is the Bar Code quality sufficient?
If you had no success, try to perform the test using the
BARCODE TEST CHART included with the product.
Communication
:
Device is not
transmitting anything to
the host
Is the serial cable connected?
Is the correct wiring respected?
Are serial host settings equivalent to the serial device
setting?
If using CBX, be sure the RS485 termination switch is
positioned to OFF.
Communication
:
Data transferred to the
host are incorrect,
corrupted or incomplete
In the Genius™ program select the Data Communication
Settings/Data Format/Standard Parameters branch and
check the Header, Separators, and Terminator values
Also check the Code Field Length and Fill Character values.
Are the COM port parameters correctly assigned?
Communication
:
Always returns the
Reader Failure
Character (<BEL> char
as default)
Contact your local Datalogic office or Datalogic Partner or
ARC, because either a Motor or Laser failure has occurred.
Note the exact model and Serial Number of the device.
How do I obtain my
units’ serial
numbers?
The device’s serial number is printed on a label that is affixed
to the body of the reader.
Serial numbers consist of 9 characters: one letter, 2
numbers, and another letter followed by 5 numbers.
Содержание DS2100N
Страница 1: ...DS2100N Reference Manual ...
Страница 20: ...DS2100N REFERENCE MANUAL 10 1 b Operating mode selection and definition c Digital Outputs configuration ...
Страница 54: ...DS2100N REFERENCE MANUAL 44 4 Figure 46 ID NET Network Connections with isolated power blocks ...
Страница 55: ...CBX ELECTRICAL CONNECTIONS 45 4 Figure 47 ID NET Network Connections with Common Power Branch Network ...
Страница 56: ...DS2100N REFERENCE MANUAL 46 4 Figure 48 ID NET Network Connections with Common Power Star Network ...
Страница 69: ...25 PIN CABLE ELECTRICAL CONNECTIONS 59 5 Figure 68 DS2100N Multidrop Connection to a Multiplexer ...
Страница 72: ...DS2100N REFERENCE MANUAL 62 5 Figure 70 ID NET Network Connections with isolated power blocks ...
Страница 73: ...25 PIN CABLE ELECTRICAL CONNECTIONS 63 5 Figure 71 ID NET Network Connections with Common Power Branch Network ...
Страница 74: ...DS2100N REFERENCE MANUAL 64 5 Figure 72 ID NET Network Connections with Common Power Star Network ...
Страница 119: ...www datalogic com ...